+8618117273997웨이 신
영어영어
中文简体 中文简体 en English ru Русский es Español pt Português tr Türkçe ar العربية de Deutsch pl Polski it Italiano fr Français ko 한국어 th ไทย vi Tiếng Việt ja 日本語

진자 충격 시험 장치

사용자 설명서

적용 할 수있는 LISUN 모델: IK01-06P

1. 장비에 대한 기본 이해

1.1 주요 매개변수

파라미터 주의 사항
Combined Mass of Impact Components 150±1 g and 200±1 g
Quality of Mounting Brackets 10 ± 1kg
Bracket Rotation Angle ± 60 °
Swing Arm Length 1000 ± 1mm
Swing Arm Quality 100g ± 1g
Material for the swing arm Steel pipe with an outer diameter of 9 mm and a wall thickness of 0.5 mm
센서 해상도 Angle: 0.36 degrees; Height: 0.1 mm
충격 에너지 0.14J, 0.2J, 0.35J, 0.5J, 0.7J, 1J

Note: According to the requirements of the IEC 60068-2-75 (GB/T 2423.55) standard and the method for calculating equivalent mass, the previous version of the standard used a 150-g impact element, whose equivalent mass was 150 g + (100 g/2) = 200 g; The new standard uses a 200 g impact element, with an equivalent mass of 200 g + (100 g/2) = 250 g. The IK01-06P includes impact elements of both masses.

See Figure 1-1 for the external dimensions.

Pendulum Impact Test Device-Figure1

그림 1-1

1.2 장비 구성 검증

1.2.1 포장 풀기

기기를 개봉할 때는 날카로운 도구로 케이스에 흠집이 나지 않도록 조심스럽게 다루십시오. 개봉 후에는 기기 외관을 먼저 검사하여 눈에 띄는 변형이나 찌그러짐이 없는지, 버튼이 손상되었거나 헐거워지지 않았는지 확인하십시오. 기기 또는 부속품에 이상이 발견되면 즉시 당사로 연락해 주십시오.

1.2.2 본체 및 부속품

Main Unit: 1 IK01-06 Impact Element Impact Test System main unit.

Main Components (Figs. 1–2):

번호 주의 사항
Expansion anchors, used to mount the main unit to a vertical wall
USB charging cable, used to charge the display
Spare 150g impact element; the 200g impact element is already installed on the swing arm at the factory.

Pendulum Impact Test Device-Figure2

그림 1~2

1.2.3 전자 문서 다운로드

제품 발송 시 디지털 사용 설명서, 보증서 및 기타 문서의 다운로드 링크를 이메일로 이미 보내드렸습니다. 가능한 한 빨리 다운로드해 주시기 바랍니다. 이메일을 받지 못하신 경우, 저희에게 연락하여 다운로드 링크를 다시 요청하실 수 있습니다.

2. 안전상의 주의

2.1 작동 안전

You must wear impact-resistant gloves and safety goggles to prevent flying debris or ricochets from the impact components from causing injury;

Unauthorized personnel are prohibited from entering the test area to prevent accidental injury.

2.2 환경 안전

The ambient temperature should be maintained between 5°C and 35°C, with relative humidity ≤ 75%;

Leave at least 2000 mm of space in front of the equipment to ensure that the swing arm can lift properly and that there are no obstructions in the surrounding area;

When mounting the equipment on a wall, ensure it is vertical, with a tolerance of ≤0.5°.

3. 장비개요

3.1 Introduction to the Host

Introduction to the Host System Components (Figure 3-1).

번호 주의 사항
Mounting holes, 12 in total
Sample holder, with adjustable angle and position
Impact element, replaceable
스윙 암
Adjustable arm mounting bracket
Fixed Device for the Impact Height of a Swing Arm
Display showing the current swing arm angle and impact height
Sensor that monitors the angle of the swing arm and automatically calculates the impact height

Pendulum Impact Test Device-Figure3

그림 3-1

3.2 Replacing The Impact Element

Steps for Replacing the Impact Element (Fig. 3-2).

번호 주의 사항
Remove the two small screws at the top and bottom
너트를 풀다
Unscrew the counterweight, then tap the connecting rod out.

Pendulum Impact Test Device-Figure4

그림 3-2

The disassembled impact components are shown in the figure below (Figure 3-3):

Pendulum Impact Test Device-Figure5

그림 3-3

Next, assemble the impact elements of different weights and install them onto the swing arm, paying attention to the orientation of the hammer head during installation.

3.3 Introduction to the Display

Display Introduction Features (Fig. 3-4).

번호 주의 사항
USB 충전 포트
Display power switch: Press and hold for 1 second to turn on; press briefly to turn off.
Confirm Button/Reset Button
The adjustment button allows you to set the length of the swing arm; the factory default setting is 1 m.
Number Up/Down keys: Increase or decrease the number when setting parameters
Battery level indicator—please recharge promptly when the battery is low.
Current Impact Height Display
Current Swing Arm Angle Display

Pendulum Impact Test Device-Figure6

그림 3-4

4. 충격시험

4.1 Sample Mounting and Installation

4.1.1 Sample Fixation

Secure the sample in the center of the plywood.

4.1.2 위치 조정

The position and angle of the sample holder are adjustable, and the position of the swing arm mounting device is adjustable. Adjust these so that the hammer head and the sample impact point are at the same height, and the direction of impact is perpendicular to the cross-section of the impact point.

4.2 Impact Parameter Settings

4.2.1 Set to 0 at Startup

When you turn on the display, it will show random angle and height values. Press the reset button to set the current swing arm angle and impact height to 0.

4.2.2 Adjusting the Impact Height

Raise the pendulum arm; the display will show its real-time angle and impact height. Refer to the table below to raise the pendulum arm to the desired impact height, and secure it using the pin on the height locking device. See Figure 4-1.

IK 레벨 IK01 IK02 IK03 IK04 IK05 IK06
에너지 J 0.14 0.2 (0.3) 0.35 (0.4) 0.5 0.7 1
Equivalent mass (kg) 0.25 (0.2) 0.25 (0.2) 0.25 (0.2) 0.25
낙하 높이(mm) 56 (100) 80 (150) 140 (200) (250) 200 280 400

Note: The values in parentheses correspond to the requirements of the previous version of the standard. If testing is to be conducted in accordance with the previous version of the standard, a 150-gram impact element must be installed.

Pendulum Impact Test Device-Figure7

그림 4-1

4.2.3 IK Impact Test

Slowly pull out the pin securing the swing arm height (Fig. 4-1) to perform the impact test.

To prevent the impact element from striking repeatedly due to inertia, after the impact element has completed an impact and bounced back, you must manually hold the impact element and slowly return it to its original position. Do not use your hand to directly grip the swing arm to stop it (as this can easily cause the swing arm to deform).

Adjust the position of the sample and continue conducting impact tests on the other test points until the entire test is complete.

4.3 사후 테스트 절차

4.3.1 Equipment Shutdown and Reset

Turn off the display and allow the impact element to hang freely.

4.3.2 Sample Preparation and Observation

Visual inspection: Check the sample housing for cracks, dents, or peeling, and check whether any fasteners are loose;

Functional Testing: When the sample is powered on, test whether all functions are operating normally and ensure there are no functional failures;

Determine whether the product meets the IK rating requirements based on the product standards, and record the testing process (test points, number of impacts, etc.) and the test results (pass/fail).

5. 장비의 일상 유지보수 및 관리

5.1 테스트 환경 준비

Clean the equipment surfaces: Wipe down the test fixture, stage, and impact elements with a dry cloth to remove dust and grease. Do not use water or corrosive cleaning agents;

Tidy up the test area: Remove any sample debris (if present), organize tools and accessories, and keep the area clean and tidy.

5.2 장기 보관 및 유지 관리 (1개월 이상 사용하지 않을 경우)

Component Storage: After cleaning the impact components and swing arms, apply rust-preventive oil and place them separately in packaging boxes to prevent collisions;

Maintenance of Lubricated Parts: The pin on the swing arm height lock and the pivot on the swing arm lock must both be lubricated to ensure proper lubrication;

Equipment Protection: Cover with a dust cover; store in a dry, well-ventilated environment at a temperature of 5–35°C and humidity of ≤75%;

Charge regularly: Store the device after fully charging it, and charge it once a month to run a self-test.

5.3 교정 주기 요구사항

측정 정확도와 장비의 작동 신뢰성을 보장하기 위해, 자격을 갖춘 제3자 계측 연구소에 정기적으로 보내 교정을 받는 것이 좋습니다. 권장 교정 주기는 12개월입니다. 사용자는 사용 빈도, 환경 조건 및 품질 시스템 요구 사항을 고려하여 구체적인 교정 주기를 결정할 수 있습니다.

6. 문제 해결 및 일반적인 문제 해결 방법

결함 증상 가능한 원인들 순서
The display won’t turn on Low battery on the display Try charging it again using the standard USB cable.
Abnormal altitude and angle readings Not reset to zero Every time the display is turned on, it shows random height and angle readings, and you must press the reset button to reset them.

7. 포장 목록

그렇지 않습니다. 이름 수량
IK01-06 Impact Element Test System Main Unit (includes a 150g impact element) 1 부
확장 나사 12
USB 충전 케이블 1 조각
200g shock element 1