RF 전도 내성 시험 시스템

사용자 설명서

적용 할 수있는 LISUN models: AB-EM

The AB-EM electromagnetic clamp complies with the national standard IEC 61000-4-6 and is a specialized product for testing the immunity of equipment to radio-frequency field-induced conducted disturbances.injecting disturbance signals into the cable assemblies in the slots so that the Equipment Under Test (EUT) is subjected to the levels specified in the national standard.

For precise measurements, the electromagnetic clamp is also equipped with an auxiliary absorption clamp and calibration pieces.

기능

See Tables 1 and 2 for the coupling coefficient and attenuation characteristic curves; the difference is 9.5 dB ± 1.5 dB.

Maximum input power: 100 W (5 minutes);

Injection frequency: 150 kHz to 230 MHz;

At frequencies greater than 10 MHz, the coupling/decoupling directivity is greater than 10 dB

Measurable cable diameter: <Φ18 mm

첨부 파일

Instruction Manual 1 copy

Test Cable One

100W 6dB Attenuator One

Calibration Fixture (Optional) One set

Using Connection Diagrams

AB-EM

RF Conducted Immunity Test System-Figure1

AB-EM Data Sheet

Serial No.: 21021 Differential Error: 9.5 dB ± 1.5 dB Date: April 29, 2021

주파수 (MHz) Coupling Coefficient (dB) Attenuation Coefficient (dB) Difference (dB)
0.15 -2.7 -12.3 9.6
0.2 -2.4 -12.0 9.6
0.3 -2.2 -11.8 9.6
0.32 -2.2 -11.8 9.6
0.35 -2.2 -11.8 9.6
0.4 -2.1 -11.8 9.7
0.45 -2.1 -11.8 9.7
0.5 -2.1 -11.8 9.7
0.55 -2.1 -11.8 9.7
0.6 -2.1 -11.8 9.7
0.65 -2.1 -11.8 9.7
0.74 -2.1 -11.8 9.7
0.75 -2.1 -11.8 9.7
0.8 -2.2 -11.9 9.7
0.85 -2.3 -11.9 9.6
0.9 -2.4 -12.1 9.7
0.95 -2.5 -12.1 9.6
1 -2.5 -12.1 9.6
1.5 -2.1 -11.7 9.6
2 -2.1 -11.7 9.6
2.5 -2.1 -11.7 9.6
3 -2.1 -11.7 9.6
4 -2.2 -11.8 9.6
5 -2.2 -11.8 9.6
6 -2.3 -11.8 9.5
7 -2.4 -11.8 9.4
8 -2.4 -11.8 9.4
9 -2.5 -11.9 9.4
10 -2.6 -11.9 9.3
12 -2.8 -12.1 9.3
15 -3.2 -12.4 9.2
18 -3.5 -12.6 9.1
20 -3.6 -12.8 9.2
22 -3.6 -12.8 9.2
25 -3.5 -12.8 9.3
28 -3.1 -12.4 9.3
30 -2.6 -11.9 9.3
32 -2.2 -11.5 9.3
35 -1.3 -10.6 9.3
38 -0.5 -9.8 9.3
40 0.0 -9.3 9.3
45 1.1 -8.4 9.5
50 1.8 -7.8 9.6
60 2.5 -7.4 9.9
70 2.6 -7.1 9.7
80 2.7 -6.9 9.6
85 2.9 -6.9 9.8
90 3.0 -6.9 9.9
95 3.2 -6.9 10.1
100 3.3 -6.8 10.1
120 3.5 -6.8 10.3
140 3.4 -7.0 10.4
160 2.6 -8.0 10.6
180 1.9 -8.2 10.1
200 0.9 -9.3 10.2
210 0.6 -9.6 10.2
220 0.1 -10.1 10.2
230 -0.1 -10.3 10.2

RF Conducted Immunity Test System-Figure2

RF Conducted Immunity Test System-Figure3

List of AB-EM Calibration Fixtures

그렇지 않습니다. 모델 이름 Quantity/piece
1 M-004 150Ω/50Ω Adapter 2
2 M-006 5 W / 50 Ω load 1
3 E-001 L-Shaped Calibration Bracket 2
4 E-003 4mm Test Rod (725mm) 1

*The 4mm test probe is inside the electromagnetic clamp packaging box;

LED Luminaires and LED Power Driver Test Solution

고니오포토미터 시스템: LSG-6000, LSG-1890B및 LSG-1700

분광복사계 및 적분구 테스트 시스템: LPCE-2 or LPCE-3

LED Life Maintains a Test System in Accordance with LM-80: LEDLM-80PL

Waterproof Test for IPX5 and IPX6 Levels: JL-56

Dustproof Testing Machine for IP5X and IP6X Testing: SC-015

LED 전원 드라이버 테스터: LS2090 (for the lab) and ATE-2 (for the production line)

Electrical Safety Tester: LS9955

CFL and Electronic Ballast Test Solution

고니오포토미터 시스템: LSG-1890B or LSG-1700

Integrating Sphere and Spectrophotometer Test System: LPCE-2 or LPCE-3

Adjustable Reference Ballast: DYJ-50HZ, HCS-109A예산 및 DYJ-HID

Electric Ballast Tester: WT5000 (for the lab) and ATE-1 (for the production line)

Digital Torsion Meter and Multiway Life Tester: CH338 CH316

전기 안전 검사원: LS9923, ZRS-3H, ZY-3

CFL 및 LED 조명기구용 EMC 및 EMI 테스트 솔루션

EMI 테스트 시스템: EMI-9KB or EMI-9KA

정전기 방전 시뮬레이터: ESD61000-2

EFT 면역력 측정: EFT61000-4

서지 생성기: SG61000-5

전압 강하 및 정전 발생기: CSS61000-11

링 웨이브 발생기: RWG61000-12

RF Conducted Immunity Test System-Figure4

WWW. Lisun그룹닷컴

특허 출원

모든 권리는 다음에 의해 보호됩니다. Lisun 그룹

사전 고지 없이 변경될 수 있습니다.