전력선용 커플링 디커플링 네트워크
사용자 설명서
적용 할 수있는 LISUN 모델: CDN61000-5P,CDN61000-5P-T440V100A,CDN61000-5P-T440V20A,CDN61000-5P-T440V32A
1. 제품 개봉 및 기본 사용법 익히기
1.1 주요 기술 사양
1.1.1 Coupling Decoupling Network For Power Lines
| LISUN 모델 | CDN61000-5P-T440V20A | CDN61000-5P-T440V32A | CDN61000-5P-T440V100A |
| Coupled Waveforms | EC61000-4-5 composite surge waveforms 1.2/50 μs and 8/20 μs, IEC61000-4-12 ringing waveform | ||
| Maximum Surge Voltage/Current | 6 kV/3 kA (Requires use with the LISUN SG61000-5 series surge generator or the RWG61000-12 Ring Wave Generator) | ||
| 상 | Asynchronous; Synchronous: 0°–360° | ||
| 커플링 방법 | Common-Mode, Differential-Mode | ||
| Capacity of the Test Equipment | Three-phase / AC 440 V / 20 A | Three-phase / AC 440 V / 32 A | Three-phase / AC 440 V / 100 A |
1.1.1 Communication Coupling Decoupling Network
| LISUN 모델 | CDN61000-5C-H8CC | CDN61000-5C-H8BC | CDN61000-5C-M8TK | CDN61000-5C-M4CC | CDN61000-5C-M4BC | CDN61000-5C-L4BK |
| 데이터 전송 속도 | ≤1000 MB/s | ≤10 MB/s | ≤100 kB/s | |||
| Capacity of the Test Equipment | 8-Channel Interconnect Cable | 4-Channel Interconnect Cable | ||||
| Coupled Surge Waveforms | 10/700 μs 및 5/320 μs | 1.2/50 μs 및 8/20 μs | 1.2/50 μs and 8/20 μs, 10/700 μs and 5/320 μs | 10/700 μs 및 5/320 μs | 1.2/50 μs 및 8/20 μs | |
| 커플 링 모드 | Common Mode | Compatible with differential-mode and common-mode signals | Common Mode | Compatible with differential-mode and common-mode signals | ||
| 포트 타입 | Unshielded, Symmetrical | Unshielded Balanced Cable, Unshielded Unbalanced Cable | Unshielded, symmetrical | Unshielded symmetrical cable, unshielded asymmetrical cable | ||
| 최대 시험 전압 | 6 kV. Note: The 1.2/50 μs CDN must be used with the SG61000-5 generator; the 10/700 μs CDN must be used with the SG61000-5C generator. | |||||
| 테스트 개체 | Ethernet and LAN (e.g., RJ45) | LIN, CAN, RS-485, RS-232, RS-422, 4–20 mA | LIN, RS-232 | |||
1.2 Unboxing Inspection and Verification of Accessories
1.2.1 포장 해제 및 육안 검사
When unpacking the device, handle it gently to avoid scratching the outer casing with sharp tools. After unpacking, first inspect the exterior of the device to ensure there are no obvious deformations or dents, and that the buttons are not damaged or loose.
If you notice any damage or deformation to the equipment or accessories, do not turn on the power. Please contact us immediately.
1.2.2 본체 및 부속품
Host Category: 1 Coupling Decoupling Network.
Standard Components for Coupling Decoupling Network for Power Lines: ① Power cord; ② RS-232 communication cable; ③ Banana plug cable; ground wire; ④ Isolation Transformer connection cable; ⑤ EUT sample connection cable; ⑥ Ground wire.

그림 1
The standard accessories for a communication line Coupling Decoupling Network consist of only two banana plug cables and one ground wire.
Download links for the digital user manual, warranty card, and other documents have been sent via email. Please download them as soon as possible. If you missed the email, please feel free to contact us to receive the download links again.
1.2.3 전자 문서 다운로드
We have already emailed you the download links for the electronic user manual, calibration certificate, warranty card, and other documents at the time of shipment. Please download them as soon as possible. If you have not received the email, please contact us to request the download links again.
2. Operation and Use
2.1 CDN61000-5P Coupling Decoupling Network For Power Lines
2.1.1 배선
Connections between the SG61000-5 Series Surge Generators and the CDN61000-5P Series Coupling Decoupling Network for Power Lines:

그림 2

그림 3
Note: For operating instructions, please refer to the SG61000-5 User Manual. Simply connect all wiring related to the built-in CDN network to the external CDN.
2.1.2 CDN61000-5C Signal Line Coupling Decoupling Network
Connections between the SG61000-5 Series Surge Generators and the CDN61000-5C Series Signal Line Coupling Decoupling Networks:

그림 4
참고: SG61000-5 series Surge Generator is used in conjunction with the CDN61000-5C series Coupling Decoupling Network, it functions equivalently to the SG61000-5C. Please refer to the SG61000-5C Operating Instructions.
3. 장비 보관
3.1 Short-Term Storage Requirements
If the equipment is used on a daily basis, no further action is required after each test and cleaning are completed.
3.2 Requirements for Long-Term Storage
If the device will not be used for an extended period (more than 1 month), please disconnect all cables and cover the device with a dust cover to prevent dust buildup.
The device should be powered on once a month for just half an hour to prevent the aging of electrical components (such as capacitors and contactors).