전압 딥 및 인터럽트 생성기
사용자 설명서
적용 할 수있는 LISUN 모델: CSS61000-11,CSS61000-11T,CSS61000-11T-100A,CSS61000-11T-32A
1. 제품 개봉 및 기본 사용법 익히기
1.1 주요 기술 사양
| LISUN 모델 | CSS61000-11 | CSS61000-11T | CSS61000-11T-32A | CSS61000-11T-100A |
| 출력 파형 | AC 사인파 | |||
| 전압 범위 | 0–110% of 220 V | 0–110% of 380 V | ||
| Internal IEC Rating | 100 %, 80 %, 70 %, 40 %, 0 % | |||
| Drop Angle Setting | 0°–360° (in 1° increments) | |||
| Duration of the voltage change | 0.5–9999.5톤 | |||
| Interval between voltage changes | 1 s–9999 ms | |||
| Number of times the voltage changes | 0-9999 번 | |||
| 작업 방법 | Fully Automatic Free-Fall Test | Select any one to three phases that drop simultaneously | ||
| 트리거 방법 | Manual, Automatic, and 50 ms | |||
| Power Supply for the EUT | Single-phase AC 220 V ±10%, max. 16 A | Three-phase AC 380 V ±10%, max. 20 A per phase | Three-phase AC 380 V ±10%, max. 32 A per phase | Three-phase AC 380 V ±10%, max. 100 A per phase |
| 계측기 전원 공급 장치 | AC 220 V (110 V optional) ±10%, 50/60 Hz | |||
1.2 Unboxing Inspection and Verification of Accessories
1.2.1 포장 해제 및 육안 검사
When unpacking the device, handle it gently to avoid scratching the casing with sharp tools. After unpacking, first inspect the exterior of the device to ensure there is no obvious deformation or dents, and that the buttons are not damaged or loose.
장비 또는 부속품에 손상이나 변형이 발견되면 전원을 켜지 마십시오. 즉시 당사에 연락해 주십시오.
1.2.2 본체 및 부속품
Main Equipment: 1 Voltage Dips and Interruptions Generator.
The CSS61000-11 single-phase EUT device looks as follows:

그림 1
The CSS61000-11T three-phase EUT device looks as follows:

그림 2
Standard Accessories: ① Equipment power cord; ② EUT (test specimen) input power supply cord; ③ EUT output connection cable; ④ Coaxial cable, which can be connected to an external oscilloscope to view waveforms; ⑤ Ground wire.

그림 3
디지털 사용 설명서, 보증서 및 기타 문서의 다운로드 링크를 이메일로 보내드렸습니다. 가능한 한 빨리 다운로드해 주시기 바랍니다. 이메일을 받지 못하신 경우, 언제든지 저희에게 연락하여 다운로드 링크를 다시 요청해 주십시오.
2. 안전상의 주의
2.1 배치 지침
Ensure the ambient temperature is between 15°C and 35°C, relative humidity between 30% and 60% (no condensation), and atmospheric pressure between 86 kPa and 106 kPa to prevent high-voltage short circuits or damage to the equipment caused by moisture;
The premises must be free of strong electromagnetic interference, corrosive gases, flammable or explosive materials, and dust, and must be well-ventilated;
Place it on a stable workbench; avoid tilting or inverting it, and keep it away from sources of vibration;
The equipment must be placed at least 1000 mm away from walls and surrounding structures (such as other testing equipment and shelving) to facilitate maintenance; at least 2000 mm of space must be left in front of the equipment to ensure smooth operation and keep the surrounding area free of obstructions.
2.2 Electrical Safety Standards
2.2.1 보안 검색
Verify that the equipment’s power supply voltage matches the supply voltage, and ensure that the power supply is properly grounded and that the grounding wire is free of damage or looseness.
Grounding Requirements: The protective grounding terminal (PG) on the main unit must be connected to ground, with a grounding resistance of ≤4 Ω.
The EUT’s ground terminal must be at the same potential as the equipment’s grounding system to ensure test safety.
2.2.2 운영 절차에 대한 안전 요구 사항
Operators must undergo professional training and be familiar with this manual and the relevant standards. Non-professionals are prohibited from operating or disassembling the equipment. The equipment contains high-voltage circuits; opening the cover may pose an electric shock hazard. Contact the manufacturer’s technical staff for repairs.
While the equipment is in operation, do not touch the EUT’s input and output wires or terminals.
Before switching connections or replacing the EUT, you must first ensure that the EUT’s power input and output switches are turned off.
여러 사람이 참여하는 경우, 역할과 책임을 명확하게 정의하고 오류를 방지하기 위해 테스트 시작 전에 서로 확인해야 합니다.
2.3 기타
본 설명서에 나와 있지 않은 사항에 대해서는 주의를 기울여 진행하시거나 당사에 문의해 주십시오.
3. 시험 전 준비 사항
3.1 Preparation of the EUT (Test Item)
Inspect the EUT to ensure there are no visible damages, no water ingress, and no rust, and that the power connector is intact.
Prepare an external power supply based on the EUT’s rated voltage and current.
3.2 Verification of Standard Parameters
3.2.1 Single-Phase Test Parameters (GB/T 17626.11)
| 테스트 레벨 | Drop Voltage (relative to UT) | 런닝타임 | 적용 가능한 시나리오 |
| 레벨 1 | 80% UT | 10 ms/20 ms/50 ms | Environment with minimal interference |
| 레벨 2 | 70% UT | 10 ms/20 ms/100 ms | General Industrial Environments |
| 레벨 3 | 40% UT | 20 ms/50 ms/200 ms | 혹독한 산업 환경 |
| 레벨 4 | 0% UT (Total Drop) | 50 ms/100 ms/500 ms | Extreme Interference Environments |
3.2.2 Three-Phase Test Parameters (GB/T 17626.34)
Drop Type: Single-phase drop, two-phase drop, three-phase drop (select according to test requirements).
Synchronization Requirements: During a three-phase drop, the initial phase difference between phases must be ≤5°, and the duration deviation must be ≤1 ms.
Recovery Requirements: After the drop ends, the voltage recovery time must be ≤ 10 ms, with no overshoot (overshoot voltage ≤ 110% UT).
3.3 Introduction to the Equipment Front Panel and Wiring
① Emergency stop switch; ② USB port, which can be used to connect a mouse or update the software; ③ The remaining three ports are reserved; if not specified at the time of purchase, they have no function;④ Touchscreen; ⑤ EUT power indicator and test indicator; ⑥ Power switch; ⑦ EUT rated voltage display; ⑧ Voltmeter displaying the EUT voltage drop; ⑨ Voltmeter displaying the EUT current voltage; ⑩ Ammeter displaying the EUT current;⑪ Connects to the EUT; ⑫ Connects to the EUT; ⑬ Connects to an external oscilloscope to view the voltage drop waveform; ⑭ Power cord; connect as specified in 2.2.1;⑮ EUT power input terminal; connect to an external power source to supply power to the EUT; ⑯ EUT power input switch; ⑰ EUT connection port; ⑱ EUT power output switch; ⑲ PG protective grounding terminal; connect as specified in 2.2.1; ⑳ EUT test specimen.

그림 4
Note: Based on the EUT’s maximum current and for ease of wiring, simply select one of ports ⑪, ⑫, or ⑰ for connection.
3.4 장비 워밍업 및 자체 테스트
Turn on the power switch on the front panel of the host; the device enters standby mode, the power indicator light turns on, and the cooling fan starts.
Wait 5 minutes for the device to warm up, and check that the touchscreen is displaying normally and that there are no error messages.
4. 터치스크린 프로그램 소개
4.1 초기화면
① General Precautions for Use (This feature is still under development; there is no need to access it at this time); ② Switch between Chinese and English; ③ System Information Screen; ④ Voltage and Frequency Dip Test Screen.

그림 5
4.2 System Information Screen
① Switch the voltage drop mode to perform the voltage drop based on either the rated voltage or the actual voltage. Example: If the rated voltage is 220 V, the actual voltage is 230 V, and the set voltage drop is 160 V, the voltage drop will be 160 V if calculated based on the rated voltage; If the voltage drop is calculated based on the actual voltage, the drop voltage will be 230/(220/160) = 167.3 V.
② Click to view the test log.

그림 6
4.3 테스트 인터페이스
① Voltage waveform display; ② Display of the EUT’s actual voltage and voltage dip; ③ EUT frequency display;④ Special function settings; ⑤ For three-phase EUTs, test parameters can be set separately for each phase (La, Lb, Lc); ⑥ Control whether to perform a voltage dip test on a specific phase; if a dip test is not required for that phase, simply set it to OFF;⑦ Rated voltage setting; ⑧ Voltage dip setting; ⑨ Voltage dip duration setting; ⑩ Interval duration setting; ⑪ Voltage dip phase angle setting; ⑫ Number of voltage dips setting; ⑬ Click to turn the EUT power on or off.

그림 7
Note 1: For the CSS61000-11 single-phase EUT Voltage Dips and Interruptions Generator, after turning on the EUT input/output switch and setting the rated voltage, the generator will automatically adjust the EUT input power supply to the set rated voltage value based on the setting. The CSS61000-11T three-phase EUT Voltage Dips and Interruptions Generator does not have an automatic rated voltage adjustment function; if you need to adjust the EUT input voltage, you must adjust the external input power supply.
Note 2: If all drop tests are to be conducted at a fixed phase angle: If your EUT sample operates at 50 Hz, set both the drop time and the interval time to be integer multiples of a single cycle (20 ms). If your EUT sample operates at 60 Hz, set both the drop time and the interval time to be integer multiples of a single cycle (16.67 ms). If the set times are not integer multiples of a single cycle, only the first dip may occur at the set phase angle.
Click to enter the special features settings screen (you may choose to enable or disable this feature; it does not affect the test).① Return to the General Settings screen; ② Automatically configure settings according to the electricity meter industry standard; each drop test consists of 3 consecutive drops (spaced 50 ms apart); ③ If your device uses a three-phase circuit for drop testing, click to have Lb and Lc automatically use the settings for La.

그림 8
5. Voltage and Frequency Dip Test
5.1 운영 절차
Turn on the EUT input power supply switch, and turn on the EUT power output switch.

그림 9
After setting all parameters on the touchscreen interface, the screen displays “Voltage Adjusting.” Observe the rated voltage and voltage dip displays on the screen, as well as the voltmeter reading. When the rated voltage (applicable only to the CSS61000-11 Single-Phase EUT Voltage Dips and Interruptions Generator) and the voltage dip are automatically adjusted to the set values, the prompt disappears.

그림 10
Note: When the above message appears, it indicates that the device is in the process of self-calibration; do not perform any other operations.
Click “EUT POWER” to connect the EUT power supply to the EUT. Follow the instructions in the EUT manual to set it to its normal operating state (e.g., power on and warm up, set the operating mode).

그림 11
Click “START” to begin the voltage dip test.

그림 12
The current voltage and current of the EUT can be monitored, and the display can be switched between the rated values and the test values.

그림 13

그림 14
참고 : CSS61000-11 series testers are capable of testing high-power EUTs; the selected current meter is a wide-range current meter. If the EUT current is relatively low, the display may not be clearly visible, but this does not affect the test.
5.2 Determining Whether the EUT Has Passed the Test
The four-level performance classification system (A/B/C/D) specified in GB/T 17626.1 (General Principles for Immunity Testing) is adopted.
| 레벨 | 정의 | Determination Result |
| PC A | During and after the test, the EUT functioned and performed completely normally, with no abnormalities observed, and complied with the technical specifications set by the manufacturer. | By |
| PC B | Temporary loss of function or temporary degradation in performance may occur during the test; however, once the disturbance ceases, the EUT shall recover to normal operation on its own without operator intervention. | Approved (subject to product specifications) |
| PC C | During testing, a loss of function or reduced performance may occur; while the system cannot recover on its own once the disturbance ceases, it can return to normal through operator intervention (such as resetting, restarting, or reconfiguring), and no permanent damage occurs. | |
| PC D | The EUT has sustained permanent damage, experienced a software failure, or suffered data loss; or its operational metrics or performance deviate from expected values and cannot be restored | 실패 |
GB/T 17626.11-2023 (Equipment ≤ 16 A) Preferred test level:
| 시험 종류 | Test Level (Percentage of Residual Voltage) | 런닝타임 | 적용 가능한 시나리오 | Recommendation Criteria |
| 전압 강하 | 40 %, 70 % | 0.01 s (half-cycle), 0.1 s, 0.5 s, 1 s | Consumer and commercial electronic devices, small industrial controllers | PC A/B (Use PC A for critical equipment) |
| Short-term interruption | 0% (Completely Interrupted) | 0.001 s (0.05 cycles), 0.01 s, 0.1 s, 1 s | All devices rated at 16A or less | PC B/C (PC B for Critical Equipment) |
| 전압 변화 | Step Changes: ±10%, ±20% | Sustained and Stable | Voltage-sensitive equipment (such as medical devices and precision instruments) | PC A |
GB/T 17626.34-2012 (Equipment >16A) Preferred Test Level:
| 시험 종류 | Test Level (Percentage of Residual Voltage) | 런닝타임 | 적용 가능한 시나리오 | Recommendation Criteria |
| 전압 강하 | 40 %, 70 % | 0.01 s, 0.1 s, 0.5 s, 1 s, 3 s | Industrial motors, large UPS systems, and power equipment | PC B/C (Use PC B for applications requiring high continuity) |
| 잠시 중단 | 0% | 0.01 s, 0.1 s, 1 s, 3 s, 10 s | Industrial automation systems, large-scale refrigeration equipment | PC C (PC C may be accepted in special circumstances) |
| 전압 변동 | Step changes: ±10%, ±20% | Sustained and Stable | Power Supplies for Large-Scale Industrial Control Systems and Data Centers | PC A/B |
6. 테스트 후 절차
6.1 Emergency Stop Procedure
6.1.1 Conditions for Triggering an Emergency Stop
The EUT presents serious safety hazards, such as smoke, fire, or explosion.
The equipment exhibits abnormal noises, odors, or overheating (casing temperature exceeds 60°C).
Loose terminal connections and short circuits caused the equipment to trip.
6.1.2 비상 대응 절차
Immediately press the emergency stop switch (red; rotate clockwise to reset), and all power to the equipment will be cut off.
If the EUT catches fire, use a dry-powder fire extinguisher to put out the fire (never use water).
Once the equipment has cooled down, check for the cause of the malfunction (e.g., wiring, parameters, EUT status).
After troubleshooting, turn the “Emergency Stop” button to reset it, and restart the equipment following the normal wiring and startup procedures.
6.2 Procedure for Concluding a Normal Test
6.2.1 Testing Complete
The test is complete, and the device has entered standby mode.
6.2.2 Power Disconnection and Disconnection of Wires from the EUT
Turn off the EUT in the following order: Click “EUT POWER” to cut off power to the EUT, turn off the EUT power output switch, and turn off the EUT power input switch.
Replace the EUT for testing, or turn off the device’s power.
6.3 Post-Test Cleanup and Organization
Clean the test area and wipe it down with a dry cloth. Do not use alcohol or corrosive cleaning agents.
6.4 장비 보관
6.4.1 단기 보관 요구 사항
장비를 매일 사용하는 경우, 각 테스트 후 청소가 완료되면 추가 조치는 필요하지 않습니다.
6.4.2 Requirements for Long-Term Storage
기기를 장기간(한 달 이상) 사용하지 않을 경우 모든 케이블을 분리하고 먼지가 쌓이는 것을 방지하기 위해 먼지 덮개로 기기를 덮어주십시오.
The device should be powered on once a month for just half an hour to prevent the aging of electrical components (such as capacitors and contactors).
7. 문제 해결 및 일반적인 문제 해결 방법
7.1 Symptom: The device does not respond after power-on (the indicator light does not come on, and the fan does not spin)
Possible causes: The power cord is not properly connected or is damaged; abnormal supply voltage; a blown fuse; a blown varistor (MOV).
Troubleshooting: Ensure the power cord is intact and securely connected. Power supply issues—such as unstable voltage, excessively high voltage, or harmonic interference—can cause the fuse or varistor to burn out. We recommend using a voltage-stabilized, pure sine wave AC power supply to power the device. Replace the fuse or varistor with one of the same type.
The figure below shows the location of the fuse:

그림 39
The figure below shows the location of the varistor (MOV) inside the filtered power supply:

그림 40
7.2 Symptom: The circuit breaker trips immediately after startup
Possible causes: Short circuit in the EUT; short circuit between the EUT input and output conductors
Troubleshooting: Check whether the EUT’s input/output wires are exposed or shorted, and replace any damaged cables; check whether the EUT is shorted (measure the resistance at the EUT’s power input terminals with a multimeter); after resolving the EUT fault, try again.
7.3 Symptoms: The “(START) Start Test” button cannot be activated
Possible causes: The voltage has not been adjusted properly; the EUT has not been powered on.
Solution: Wait 3 to 5 seconds to allow the device to automatically adjust the voltage to the set value (wait until the Vr and Vd readings stabilize); click the “EUT Power” button to power on the EUT, then try again.