디지털 CC 및 CV DC 전원 공급 장치

사용자 설명서

적용 할 수있는 LISUN 모델: DC3005S,DC3010S,DC6005S

1. 제품 개봉 및 기본 사용법 익히기

1.1 주요 기술 사양

사양/모델 번호 DC3005S DC3010S DC6005S
최대 출력 30 V / 5 A 30 V / 10 A 60 V / 5 A
전압 분해능 0.001 V (0–60 V)
전압 정확도 ±(0.02% of reading + 0.05% of range + 1 digit)
현재 해상도 0.001
현재 정확도 ±(0.05% of reading + 0.05% of full scale + 1 digit)

1.2 Equipment Configuration Verification

1.2.1 포장 해제 및 육안 검사

When unpacking the device, handle it gently to avoid scratching the outer casing with sharp tools. After unpacking, first inspect the exterior of the device to ensure there is no visible deformation or dents, and that the buttons are not damaged or loose. If you notice any abnormalities with the device or its accessories, do not turn it on. Please contact us immediately.

1.2.2 본체 및 부속품

번호 주의 사항
DC-S Series Power Supplies
전원 코드
RS-232 통신 케이블

Download links for the electronic user manual, software warranty card, and other documents have been sent via email. Please download them as soon as possible. If you missed the email, please feel free to contact us to request the download links again.

참고 : 모든 LISUN spectroradiometer software can control the DC-S series power supplies; however, the DC-S power supplies sold separately do not come with dedicated software. Please operate them using the front panel of the device.

2. 안전상의 주의

2.1 Electrical Safety Standards

The equipment’s input power supply is single-phase AAC 220 V ±10%, 50/60 Hz. It must comply with safety standards and have a reliable protective grounding system with a protective grounding resistance of ≤4 Ω to prevent electric shock.

2.2 Environmental Safety Standards

For indoor use only; avoid direct sunlight. Place on a sturdy, stable surface to prevent excessive vibration. Ensure the environment is free of dust, corrosive gases, and flammable or explosive materials. Keep away from strong magnetic and electric fields. Ensure adequate ventilation; leave at least 10 cm of space around the device for heat dissipation, and do not block the cooling fan vents. Operating Environment: Temperature 5°C to 35°C, relative humidity ≤65% RH.

2.3 Operational Safety Guidelines

Do not operate the equipment or plug in or unplug the power cord with wet hands; Output must be turned off before connecting or disconnecting wires; do not touch the output terminals, internal components, the load, or load wiring while the equipment is operating; do not shut down the equipment directly while it is in output mode; turn off the output first before shutting down the equipment; if you notice an unusual odor, smoke, unusual noise, or overheating, immediately disconnect the power and do not continue operation.

3. 장비 개요 및 배선

3.1 Introduction to the Front Panel

번호 주의 사항
전원 스위치
Voltage display window V; “—–” appears when the voltage exceeds the limit.
Current display window A; “—–” appears when the limit is exceeded.
Indicator lights. CC: Constant Current; CV: Constant Voltage; OVP: Overvoltage Protection Triggered; OCP: Overcurrent Protection Triggered
Memory Keys: The device can store 5 sets of configuration data, allowing for quick setup.
⑥⑦ You can use the number keys or the up and down arrows to adjust the settings.
V-Set: Press to set the voltage
I-Set: Press to set the current
OVP set—press to set the upper limit of the output voltage
OCP set—press to set the upper limit for the output current
After setting the values, press the ENTER key to confirm.
ON/OFF button: Tap to turn the power output on or off
Output terminals: Connect to your DC load, observing the correct polarity.

Digital CC And CV DC Power Supply-Figure1

3.2 Introduction to the Back Panel

번호 주의 사항
Power cord receptacle, with a fuse, for connecting the device to an appropriate power source
RS-232 communication port: If you need to use other software to configure and control the power output, please connect it to a computer.
Output terminals; these have the same function as the output terminals on the device’s front panel
The external sampling terminal, which is also connected to your DC load, allows you to obtain a more accurate load-side voltage. If you require high voltage sampling accuracy, please connect it; leaving it unconnected does not affect the DC power supply’s output functionality. Be sure to observe the correct polarity when wiring.
Sampling Selector Switch The factory default is Side A, with internal sampling.
If you have an external sampling terminal connected, switch to Side B for external sampling.

Digital CC And CV DC Power Supply-Figure2

Note 1: External sampling uses a 4-wire configuration to eliminate cable voltage drop; internal sampling includes cable voltage drop.

Note 2: For the output terminals, connect either the front panel or the rear panel—one or the other is sufficient.

4. 장비 작동

4.1 Verification of OVP/OCP Settings

To ensure normal operation, the device is set to its maximum output value at the factory. For example, for the DC3005S, the OVP value is set to 30 and the OCP value is set to 5. If you need to review and adjust these settings, follow the steps below.

명령 주의 사항
Press the OVP Set button; the voltage display will show the currently set OVP value. If it is incorrect, you can use the numeric keys to reset it, then press the ENTER key to confirm.
Press the OCP Set button; the current display will show the currently set OCP value. If it is incorrect, you can use the numeric keys to reset it, then press the ENTER key to confirm.

4.2 Constant-Voltage Output

명령 주의 사항
If you want the device to save the settings, press the corresponding memory button first.
Press the I-Set key to set the current to the maximum output value, or at least to a value sufficient for constant-voltage output. Use the numeric keys to set the value directly, then press the ENTER key to confirm.
Press the V-Set key to set the voltage to the desired rated value, then press the ENTER key to confirm.
Press the ON/OFF button to control the output; the CV indicator lights up when the output is active.

Note: If, during output, the CV indicator turns off and the CC indicator turns on, the set current value is insufficient to maintain the rated voltage of the DC load; you will need a DC power supply with a higher maximum output current.

4.3 Constant-Current Output

명령 주의 사항
If you want the device to save the settings, press the corresponding memory button first.
Press the V-Set key to set the voltage to the maximum output value, or at least to a value sufficient for constant-current output. Use the numeric keys to set the value directly, then press the ENTER key to confirm.
Press the I-Set button to set the current to the desired rated value, then press the ENTER button to confirm.
Press the ON/OFF button to control the output; the CC indicator light turns on when the output is active.

Note: If, during debugging, the CC indicator turns off and the CV indicator turns on, this indicates that the set voltage value is insufficient to meet the rated current output of the DC load. You will need a DC power supply with a higher maximum output voltage.

5. 장비의 일상 유지보수 및 관리

5.1 일상적인 유지보수

5.1.1 일일 청소

Be sure to turn off the power and unplug the power cord before cleaning. We recommend wiping the unit with a soft, antistatic cotton cloth and a mild detergent; do not spray liquids directly onto the unit. Do not use chemical solvents such as benzene, toluene, or acetone.

5.1.2 정기 검사

Before each use, check that the power cord and terminal connections are secure and free of burn marks; ensure that the fan vents are not blocked; and after turning on the unit, verify that the indicator lights and display are functioning properly.

5.2 Storage Specifications

5.2.1 저장 환경 요구 사항

Temperature: 5°C to 35°C; relative humidity: ≤65% RH; dry, dust-free environment free of corrosive gases.

5.2.2 장기 저장 작업

Turn off the power and unplug all cables; clean the equipment and wrap it in dust-proof film; turn on the power for 10 minutes every 3 months to prevent moisture damage to the components.

문제 해결 및 일반적인 문제 해결 방법

결함 증상 가능한 원인들 순서
전원이 켜지지 않음 Loose input power supply cord, blown fuse Turn off the power and check the connection of the input power supply cord; turn off the power, troubleshoot for short circuits or overloads, replace the fuse with one of the same rating, and do not increase the current rating.
The settings are fine, but the OCP and OVP indicator lights are on. Output Short Circuit, Output Out-of-Limits, and Device Overheating Protection Stop output, check the output wiring, and retry after ruling out a short circuit; verify that the load’s rated voltage and current are within limits—for example, the maximum output voltage and current for the DC3005S model are 30V and 5A; shut down the unit, allow it to ventilate, clean the ventilation openings, and restart it after it has cooled down.
The fan isn’t spinning Dust buildup, motor failure Turn off the power and clean the dust from the fan; if the problem persists, please contact us for repair or to have the fan replaced.

7. 포장 목록

그렇지 않습니다. 이름 수량
DC-S Series Power Supply Unit 1 부
전원 코드 1 조각
RS-232 통신 케이블