고온 및 저온 습도 챔버

사용자 설명서

적용 할 수있는 LISUN models: GDJS-010A、GDJS-010B、GDJS-010C、GDJS-010D、GDJS-013A、GDJS-013B、GDJS-013C、GDJS-013D、GDJS-015A、GDJS-015AS、GDJS-015B、GDJS-015C、GDJS-015D、GDJS-020A、GDJS-020B、GDJS-020C、GDJS-020D、GDJS-030A、GDJS-030B、GDJS-030D、GDJS-100A、GDJS-100B、GDJS-100C、GDJS-100D、GDJS-225A、GDJS-225B、GDJS-225C、GDJS-225D、GDJS-500A、GDJS-500B、GDJS-500C、GDJS-500D、GDJS-800A、GDJS-800B、GDJS-800C、GDJS-800D

1. 장비에 대한 기본 이해

1.1 주요 기술 사양

LISUN 모델 GDJS-100 * GDJS-225 * GDJS-010 * GDJS-015 * GDJS-020 *
Working Compartment Dimensions (cm) 41 × 55 × 45 50 × 60 × 75 70 × 80 × 90 100 × 100 × 100 100 × 100 × 150 100 × 100 × 200
외부 치수(cm) 105 × 105 × 156 105 × 102 × 200 132 × 132 × 217 167 × 152 × 231 245 × 160 × 231 160 × 254 × 214
작동 전력 5.5kW 7.0kW 13.5kW 15.0kW 16.5kW 26.5kW
온도 범위 A: -20°C to 150°C B: -40°C to 150°C C: -60°C to 150°C D: -70°C to 150°C
Variability/Uniformity ±0.5°C/±2°C
Rate of temperature rise 분당 1.0°C~3.0°C
냉각 속도 분당 0.7°C~1.0°C
습도 범위 20%–98% 상대습도
습도 편차 -2 %에서 -3 %

1.2 개봉 및 내용물 확인

1.2.1 포장 해제 및 육안 검사

Handle the unit with care when unpacking it. Do not use sharp tools to avoid scratching the equipment’s exterior. After unpacking, inspect the main unit first: check that the housing shows no deformation, dents, or peeling paint, and ensure that the control panel instruments and buttons are not damaged or loose. If you discover any damage to the device or its accessories, do not power it on. Contact LISUN 고객 서비스 담당자가 즉시 응대합니다.

1.2.2 본체 및 부속품

1 GDJS Series High and Low Temperature Humidity Chamber.

The spare parts box is typically mounted inside the test chamber. Please open the spare parts box; standard spare parts are shown in Figure 1-1.

번호 주의 사항
배수관
Rubber plug, used to seal the cable openings on the side of the test chamber
Spare heating elements: If the equipment has been in use for a long time and the heating elements inside the boiler are damaged, users can replace them themselves.
Spare filters; the test chamber requires regular filter replacement.
Spare fuses, spare
Backup Solid-State Relay

High And Low Temperature Humidity Chamber-Figure1

그림 1-1

1.2.3 전자 문서 다운로드

We have already sent download links for the electronic user manual, calibration certificate, warranty card, and other documents via email at the time of shipment. Please download them as soon as possible. If you have not received the email, please contact us to request the download links again.

1.3 장비 구조 설명

1.3.1 시험 챔버의 정면도

그림 1~2을 참조하십시오.

번호 주의 사항
제어 패널
The test chamber door is either a single door or a double door, depending on the size of the test chamber.
Condenser Vent

High And Low Temperature Humidity Chamber-Figure2

그림 1~2

1.3.2 시험실 후면

그림 1~3을 참조하십시오.

번호 주의 사항
냉매 압력 게이지
This is an overflow outlet; a small amount of water will flow out during the test. Simply connect this water pipe to a drain. A water pipe with an inner diameter of 10 mm is required.
This valve serves as a drain outlet. Open it when the equipment has not been used for an extended period or when the boiler needs to be cleaned to drain the water from inside the equipment. A water pipe with an inner diameter of 8 mm is required.
Water Inlet: The test chamber requires water to conduct humidity tests. Since the test chamber does not have a water pump, it must be connected to a clean water source with a pressure of at least 0.3 MPa (preferably purified water free of impurities and minerals, as otherwise the heating elements inside the boiler may burn out). Otherwise, water cannot enter the boiler, and the test chamber will be unable to perform humidity tests. The water tank can be placed in a safe location on top of the test chamber, utilizing the difference in elevation to generate water pressure. A water pipe with an outer diameter of 8 mm is required.
전원 코드

High And Low Temperature Humidity Chamber-Figure3

그림 1~3

1.3.3 Side of the Test Chamber

High And Low Temperature Humidity Chamber-Figure4

그림 1~4

1.3.4 Inside the Test Chamber

그림 1~5을 참조하십시오.

번호 주의 사항
Digital Temperature and Humidity Gauge
환기 덕트
Shelves—the number depends on the size of the test chamber; the shelf height is adjustable; samples must be placed on the shelves.

High And Low Temperature Humidity Chamber-Figure5

그림 1~5

2. 안전상의 주의

2.1 전기 안전

2.1.1 Power Connection Specifications

The equipment must be powered by a dedicated circuit with a three-phase, five-wire system rated at AC 380 V, 50 Hz (or 60 Hz, optional). Multiple units must not share a single power source to prevent voltage drops that could cause abnormal performance or equipment failure. Before connecting the power supply, ensure the power is turned off. When wiring, strictly distinguish between the live wires (L1, L2, L3), neutral wire (N), and ground wire (PE). The wiring sequence must comply with the requirements of the equipment’s electrical schematic; reversing the live and neutral wires is strictly prohibited.

High And Low Temperature Humidity Chamber-Figure6

그림 2-1

After completing the wiring, pull firmly on the terminal blocks to ensure the connections are secure and free of looseness.

2.1.2 접지 요구 사항

The equipment must be reliably grounded, with a grounding resistance of ≤4 Ω. Do not use gas pipes, non-metallic water pipes, or neutral wires as substitutes for grounding wires. After grounding is complete, use a Grounding Resistance Tester to measure the grounding resistance. Do not energize the equipment until you have confirmed that it meets the requirements, to prevent electric shock, equipment malfunctions, or measurement errors caused by poor grounding.

2.1.3 Electrical Operation Precautions

Before wiring, inspecting, or maintaining the equipment, you must turn off the circuit breaker on the back of the unit and disconnect the main power supply, and confirm that there is no residual voltage before proceeding;

감전 위험을 방지하기 위해 젖은 손으로 장비의 전원 스위치, 터치스크린 또는 계기판 버튼을 조작하지 마십시오.

전기 제어 캐비닛 내부에는 먼지나 이물질이 들어가 단락을 일으키거나 부품을 손상시킬 수 있으므로 공구나 종이 등의 이물질을 넣지 마십시오.

Do not plug in or unplug power cords or signal cables while the equipment is running to prevent short circuits that could damage the controller or electrical components.

2.2 작동 안전

2.2.1 Test Specimen Requirements

The equipment must not be used for testing flammable, explosive, toxic, or highly corrosive substances (such as alcohol, gasoline, strong acids, strong bases, etc.). The equipment is not equipped with explosion-proof devices, and such substances may cause fires, explosions, or corrosion of the equipment;

If the test specimen generates heat (such as during powered testing of electronic components), an external, independent power source must be provided for the specimen. Do not use the equipment’s own power supply directly to avoid overloading the equipment’s circuitry;

The sample volume must not exceed 80% of the test chamber’s capacity, and the weight must not exceed the specified load capacity of the shelves (30 kilograms per shelf) to prevent deformation of the shelves or disruption of the temperature and humidity cycling.

2.2.2 작동 안전

Non-professionals are prohibited from disassembling or repairing the equipment while it is in operation. In the event of a malfunction, contact LISUN Customer Service or follow the instructions of a technician.

Avoid opening the chamber door as much as possible when temperatures are high (≥80°C) or low (≤-20°C): Opening the door may cause burns or frostbite and disrupt the temperature and humidity balance (affecting test results).

If you must open the door, wear heat-resistant gloves and safety goggles, open the door slowly, and wait until the internal temperature and humidity have stabilized before proceeding.

Do not set the temperature or humidity above the device’s operating range; doing so may burn out the heating element or damage the cooling system.

If the equipment triggers an alarm, stop the test immediately and troubleshoot the problem (see Chapter 7). Do not force the equipment to restart (to prevent further damage).

2.2.3 금지된 행위

Do not open the control cabinet door while the equipment is running to avoid touching the high-voltage components inside and causing an electric shock;

Do not disassemble core components such as the temperature sensor (PT100 platinum resistance thermometer), heating element, or condenser without authorization. Modifying the equipment will void the warranty and may result in safety incidents;

Do not place heavy objects (such as tools or accessories) on top of the device to avoid damaging the device panel or impairing heat dissipation;

Do not pile flammable items (such as cardboard boxes, plastic wrap, etc.) around the equipment; keep the equipment’s ventilation openings clear to prevent the risk of fire.

2.3 환경 안전

2.3.1 설치 요구 사항

The equipment must be installed in an air-conditioned room, with the ambient temperature maintained between 5°C and 25°C and relative humidity ≤85%. If the ambient humidity is high, condensation may form near the equipment door and test ports; in such cases, improve room ventilation or use a dehumidifier;

The installation site must be free of high concentrations of dust, corrosive gases (such as sulfur dioxide and chlorine), and flammable or explosive atmospheres. Direct sunlight must not be allowed to shine on the equipment to prevent corrosion of the equipment housing and deterioration of electrical components;

Keep the device away from areas with significant fluctuations in ambient temperature, such as near vents or room entrances, to prevent the device’s temperature control accuracy from being affected.

2.3.2 공간 요구 사항

The distance between the equipment enclosure and the wall, as well as surrounding structures (such as other test equipment and shelves), must be at least 1500 mm to ensure unobstructed ventilation at the equipment’s heat dissipation openings (rear and sides);

At least 2000 mm of space must be left at the equipment door to ensure that the door can be opened in any direction, facilitating the placement and removal of samples;

2.3.3 Horizontal Placement

The equipment must be placed on a level surface. Use a spirit level to ensure the equipment is level: Place the spirit level on the top surface of the equipment and confirm that the bubble is centered. Note: The following types of legs are adjustable to ensure the test chamber is level; see Figure 2-2:

High And Low Temperature Humidity Chamber-Figure7

그림 2-2

If the equipment is tilted, it may compromise the structural stability of the test chamber, affect the liquid levels in all internal water containers, cause uneven distribution of temperature and humidity within the chamber, and even result in an uneven chamber door that does not seal properly, thereby affecting test results and the equipment’s service life.

2.4 기타

본 설명서에 나와 있지 않은 사항에 대해서는 주의를 기울여 진행하시거나 당사에 문의해 주십시오.

3. 시험 전 준비 사항

3.1 Connecting Water Pipes and Power Cables

See sections 1.3.2 and 2.1.

3.2 샘플 배치

For sample restrictions, see Section 2.2.1. For sample placement, see Sections 1.3.3 and 1.3.4.

Note: If the sample requires the use of a cable, please reseal the side-opening channel on the outside of the test chamber using a rubber plug, as shown in Figure 3-1:

High And Low Temperature Humidity Chamber-Figure8

그림 3-1

Samples must not come into contact with the inner walls of the test chamber; do not touch or cover the temperature and humidity probes; and do not block the air ducts. Samples must be securely fastened and must not move during the test.

3.3 Closing the Cabinet Door

Make sure the box door is closed and the lock is secured. For double doors, first press and lock the left door, then press and lock the right door.

3.4 Starting Up the Test Chamber

3.4.1 Turn on the power to the test chamber

Press the “Power” button on the control panel on the front of the device. The touchscreen program will enter the standby screen. Ensure that there are no alarms in the program; if there are any, refer to Chapter 7 for troubleshooting.

3.4.2 Humidity System Ready

After startup, the inlet water solenoid valve opens automatically, and water is automatically fed into the boiler. You must wait until the boiler is sufficiently filled with water to completely submerge the humidification tube before performing a humidity test. If you only need to perform a temperature test and do not need to perform a humidity test, you may skip this step.

High And Low Temperature Humidity Chamber-Figure9

그림 3-2

During the humidity test, the water in the boiler is gradually consumed, and the water inlet solenoid valve automatically opens to replenish the boiler with purified water.

4. Touchscreen Program Operation Of The High And Low Temperature Humidity Chamber

4.1 Introduction to the Main Screen

High And Low Temperature Humidity Chamber-Figure10

그림 4-1

High And Low Temperature Humidity Chamber-Figure11

그림 4-2

4.2 Operation Settings

Go to the “Operation Settings” screen, as shown in Figure 4-3.

번호 주의 사항
Select the test operation mode: FIX (fixed values) or PRO (program)
Once the test is complete, no action will be taken or the system will return to room temperature. If you want the system to automatically return to room temperature, please set the room temperature range.

High And Low Temperature Humidity Chamber-Figure12

그림 4-3

4.3 Appointment Settings

Go to the “Scheduled Settings” screen, as shown in Figure 4-4. The left column displays the current time, and the right column allows you to enter the scheduled time. Then, press the “RESERVE” button on the right until it turns red to confirm the schedule. The test chamber will then automatically begin the impact test at the scheduled time.

High And Low Temperature Humidity Chamber-Figure13

그림 4-4

4.4 디스플레이 설정

Go to the “Display Settings” screen, as shown in Figures 4-5.

번호 주의 사항
Select Display Language
Restart Methods Following an Unexpected Power Outage (Stop: An operation that restores the system to a “stop” state when power is restored after a power outage during operation; Cold Start: An operation that restarts the system from the first section when power is restored after a power outage during operation; Hot Start: An operation in which, following a power outage during operation, operation continues from the section where it was interrupted prior to the outage, restoring the system to its pre-outage state when power is restored)

High And Low Temperature Humidity Chamber-Figure14

그림 4~5

4.5 Fixed-Value Operation (Single Temperature and Humidity Point Test)

4.5.1 Accessing the Fixed-Value Settings

Select “Fixed Value” in “Operation Settings.” Go to the “Monitoring Screen,” then click the “Settings” button on the right to access the “Fixed Value Settings” screen, as shown in Figure 4-6.

High And Low Temperature Humidity Chamber-Figure15

그림 4~6

4.5.2 매개변수 설정

그림 4~7을 참조하십시오.

번호 주의 사항
① Temperature Setting Enter the target temperature (see Section 1.1 for the configurable temperature range)
② Humidity Setting Enter the target humidity (e.g., 60% RH); the range is 20%–98% RH. Conventional High and Low Temperature Chambers allow you to set the humidity range and the corresponding temperature range; see Figure 4-8.
③ Scheduled Operations Enter the test duration (e.g., 2H; range: 0–9999H); if set to 0, the test must be stopped manually.
④ Wait for the settings to take effect If the “wait” function is selected, the test chamber will not begin counting the test time until it reaches the set temperature and humidity. You must also set the tolerance range (the test is considered complete if the values fall within this range) and the wait time (if the test chamber has not reached the set temperature and humidity by the end of the wait time, it will automatically begin counting the test time).

High And Low Temperature Humidity Chamber-Figure16

그림 4-7

High And Low Temperature Humidity Chamber-Figure17

그림 4-8

4.5.3 Starting the Calibration Test

After verifying that the parameters are correct, return to the monitoring screen and click the “Run” button in the lower-right corner.

Note: If you have set both temperature and humidity, the test chamber will first adjust the temperature to the target value; once the temperature reaches the target value, the humidity function will begin to operate. Until the temperature stabilizes, the humidity function will not operate and will remain unstable.

4.6 Test Procedure (High- and Low-Temperature and Humidity Cycling Test)

4.6.1 Program Group Settings Screen

그림 4~9을 참조하십시오.

번호 주의 사항
Program number: You can set up to 100 programs, numbered 1 through 100, to avoid duplicating existing programs.
Enter the name of this program
프로그램 편집
Loop Settings
Waiting for configuration

High And Low Temperature Humidity Chamber-Figure18

그림 4-9

4.6.2 Accessing the Program Editing Screen

그림 4-10를 참조하세요.

번호 주의 사항
Test segment number; each program can be set to have up to 99 test segments
온도 설정
습도 설정
Test Duration Settings
The relay operates automatically; no parameter changes are required here.
Toggle Section Numbering
Enter the number of sections you want to run. For example, if you only want to run the first four sections, enter 4. Please enter the correct number; otherwise, the program will encounter an error.
Return to the Program Group Settings Screen

High And Low Temperature Humidity Chamber-Figure19

그림 4-10

Example: If you need to perform the following test: Step
1: Temperature -20°, no humidity, test duration 3 hours;
Step 2: Temperature 50°, humidity 50%, test duration 5 hours. The

settings should be:

NO. 01 02 03 04 05 ...
온도 -20 ° 50 ° 0
습기 0 50
Time 0.01.00 3.00.00 0.01.00 5.00.00

NO 01: Transition phase, i.e., the time required to reach the target temperature.
NO 02: Actual test duration: 3 hours.
NO 03: Transition phase—the time required to reach the target temperature and humidity.
NO 04: Actual test duration—5 hours.

In other words, for each set of temperature and humidity conditions, two phases must be configured: the first is the transition phase, during which the test chamber reaches the set temperature and humidity; the second is the test duration at those conditions.

Note: In the examples above, the transition phase is set to only 1 minute. The equipment has a wait function; even if the set time is short, the equipment will reach the set values before proceeding to the next phase. This configuration minimizes the equipment’s preparation time to save on test time.

4.6.3 Loop Settings Screen

그림 4-11를 참조하세요.

번호 주의 사항
The number of times the program loops; the default value is 1, meaning the program stops immediately after the current test is completed.
Link program number: After the current program’s test is complete, the system will automatically proceed to the next program to continue testing. If this is not desired, enter 0.
In the current program’s test segment loop mode, the default value for all parameters is 0, meaning the test segments are not repeated; instead, the program tests the segments sequentially from the first to the last.

High And Low Temperature Humidity Chamber-Figure20

그림 4-11

Example: Figure 4-12 means that after the current program has been run twice, it automatically proceeds to Test Program 3 to continue testing.

High And Low Temperature Humidity Chamber-Figure21

그림 4-12

Figure 4-13 illustrates that the current test sequence for the program is 01>02>03>04>05>03>04>05>02>03>04>02>03>04>05>06>07…
meaning testing begins with the first segment.sections 3 through 5 are repeated twice, followed by sections 2 through 4 being repeated twice, and then the remaining test segments are completed in sequence.

High And Low Temperature Humidity Chamber-Figure22

그림 4-13

Based on the logic above, you can configure the program’s test parameters to meet your specific testing requirements.

4.6.4 Wait for the Settings Screen

그림 4-14를 참조하세요.

① Select “ALL” to enable the wait function for all test segments; ② Error tolerance for temperature and humidity wait zones; ③ Maximum wait time.

High And Low Temperature Humidity Chamber-Figure23

그림 4-14

4.6.5 Startup Test

After verifying that the parameters are correct, return to the monitoring screen and click the “Run” button in the lower-right corner.

4.7 Test Operation and Monitoring

4.7.1 Monitoring Screen

The monitor screen displays the set temperature and humidity values, as well as the current actual temperature and humidity.

4.7.2 실시간 고장 모니터링

If a fault occurs, the touchscreen automatically switches to the “Real-Time Fault Display Screen.”

Check the error codes and refer to Chapter 7 for troubleshooting and resolution; click “Mute” to turn off the buzzer, and click “Reset Errors” after troubleshooting is complete.

4.7.3 Viewing and Exporting Curves

Tap the “Curve Display” screen on the main screen to view real-time temperature and humidity curves (temperature: red; humidity: blue).

Insert the USB flash drive, click “Download,” and the curve data will be saved to the USB flash drive in CSV format (which can be opened and analyzed on a computer).

Tap “Clear History” (press and hold for 5 seconds) to delete old curve data (use with caution).

5. 테스트 후 절차

5.1 Equipment Shutdown Procedures

5.1.1 Program Termination

If the program ends normally: The device operates according to the “Program End Mode,” such as “Return to Room Temperature.”

To stop the process early: Click the “Stop” button on the monitoring screen, and the device will enter standby mode.

5.1.2 전원 끄기 절차

Verify that the device is in standby mode (not running), then turn off the power switch on the device’s control panel.

주 전원 스위치를 끕니다.

5.2 Sample Retrieval and Handling

Wear heat-resistant gloves (as the sample may still be hot or cold even after returning to room temperature), open the chamber door, and slowly remove the sample.

Inspect the condition of the samples (e.g., appearance, performance) and record test data (temperature, humidity, time, changes in the samples).

If there is any liquid remaining on the sample, wipe it clean with a lint-free cloth to prevent contamination of the work area.

5.3 장비 청소

Cleaning the Inner Chamber: Wipe the inner walls and shelves of the chamber with a lint-free cloth dampened with alcohol to remove any residual sample stains (do not spray water or alcohol directly onto the surface).

Cleaning the Refrigerator Door: Wipe down the door gasket (to remove dust) and check that it is not deformed (if there are stains, wipe them off with a damp cloth).

Cleaning the Exterior: Wipe down the equipment casing with a dry cloth to remove surface dust (do not use gasoline or organic solvents, as this may cause the paint to peel).

Cleaning the Touchscreen: Wipe with a dry cloth to remove fingerprints and ensure touch sensitivity.

5.4 Test Records

Fill out the equipment usage log, recording the following information: test date, program group, sample name, runtime, fault details (if any), and operator.

Organize test data (such as graphs and sample reports) and file them for safekeeping.

6. 장비의 일상 유지보수 및 관리

6.1 Routine Cleaning and Maintenance (Before and After Each Test)

유지 관리 프로젝트 유지 보수 절차 중요 사항
Inner Chamber of the Chamber Wipe with an alcohol-soaked lint-free cloth to remove stains Do not wash with water; avoid exposing electrical components to moisture.
씰링 스트립 Wipe with a dry cloth and check for any deformation or damage. If the rubber seal is worn out (and no longer seals properly), contact customer service to have it replaced.
터치 스크린 Wipe with a dry cloth to remove fingerprints and dust Do not scratch the screen with sharp objects.
Vent (Air-Cooled) Wipe off surface dust with a dry cloth and ensure proper ventilation. Do not block the ventilation openings (to prevent overheating)

6.2 Regular Inspection and Maintenance (Performed on a Scheduled Basis)

유지 관리 프로젝트 유지보수 주기 정비 지침 목표 및 요구 사항
배관 시스템 2 개월마다 1. Clean the water filter cartridge; 2. Drain the water tank and rinse it out The water lines are clear of blockages, and there are no leaks at the connections.
온도 및 습도 센서 월간 회원 Use a soft-bristled brush to remove dust from the sensor surface Measurement accuracy within ±2°C/±3% RH
Condenser (Air-Cooled) 3 개월마다 Use compressed air (0.2 MPa) to blow dust off the condenser Good heat dissipation, with no decrease in cooling efficiency
접지 저항 Grounding Resistance Tester Inspection 접지 저항 ≤ 50Ω
전기 구성 요소 6 개월마다 Open the control cabinet door and check that the relays and terminal blocks are secure. No oxidation or loose connections; the circuit breaker is functioning normally.

6.3 Long-Term Downtime for Maintenance (More Than 1 Month)

6.3.1 Preparations Before Shutdown

Complete the final cleaning (inner tank, outer casing, and ventilation openings) to ensure no stains remain.

Drain the water from the tank and water lines (to prevent freezing and expansion from damaging components).

6.3.2 Power Outage and Protection

Turn off all power sources (device power switches, main power supply), disconnect power cords and water lines, and organize the cables (to prevent them from being pinched).

Cover the equipment with a plastic dust cover and store it in a dry room (temperature 5–25°C, humidity ≤ 60%).

6.3.3 정기점검

Every 15 days, open the dust cover, ventilate for 30 minutes, and check the inner liner for moisture (if there is any moisture, wipe it off with a dry cloth).

Start the equipment once a month, set it to 50°C, and run it for 30 minutes (no load) to dry out any internal moisture and extend the life of the components.

7. 문제 해결 및 일반적인 문제 해결 방법

7.1 On-Screen Error Codes and Troubleshooting

문제 코드 가능한 원인들 해법
X2 Compressor No. 1 Overcurrent 1. Thermal overload protection has tripped; 2. Excess refrigerant 1. Press the white reset button on the thermal overload protector; 2. Use a multifunction gauge to check the pressure in the low-pressure line; if the pressure is too high, release some refrigerant.
X3 Compressor #1 Overpressure 1. Poor condenser cooling; 2. Excess refrigerant; 3. Water-cooled unit is not turned on 1. Air-cooled: Clean dust from the condenser; Water-cooled: Turn on the water-cooling unit; 2. Check the refrigerant pressure and release any excess refrigerant
X10 Test Chamber Over-Temperature 1. Over-temperature protection triggered; 2. Set temperature exceeds the upper limit; 3. Dry burning of the heating element 1. Reset the over-temperature protection dial by turning it; 2. Reset the temperature; 3. Check the water lines to ensure the humidification system is supplying water properly
PT1 Temperature Probe Failure 1. Loose PT100 sensor wiring; 2. Sensor failure 1. Open the control cabinet door, check the PT1 wiring, and secure it again; 2. Use a multimeter to measure the sensor resistance (approximately 100 Ω at room temperature); replace it if the reading is abnormal.
PT2 Humidity Sensor Malfunction 1. Loose connection in the humidity sensor; 2. Sensor damaged by moisture 1. Check the PT2 wiring and secure it properly; 2. Replace the humidity sensor and ensure there are no obstructions at the installation site.

7.2 Controller-Related Failures

결함 증상 가능한 원인들 해법
The controller has no digital display 1. The power cord is not plugged in properly; 2. A phase is missing in the three-phase power supply; 3. Controller malfunction 1. Check the power cord and circuit breaker to ensure power is connected; 2. Use a multimeter to measure the three-phase voltages (all around 380V); 3. Contact customer service to have the controller replaced.
The numbers on the controller display are fluctuating erratically 1. Loose sensor connections; 2. Significant power supply interference; 3. Controller malfunction 1. Unplug and replug the sensor cable; 2. Move the device away from high-power equipment, or install a power filter; 3. Contact customer service for repair.
PLC Communication Error (Displays “PLC no response”) 1. The communication cable between the controller and the PLC is loose; 2. The PLC power supply is malfunctioning 1. Check the communication cable (RS485), and either secure it or replace it; 2. Check the PLC power supply (24V DC); if there is an abnormality, repair the power supply module.

7.3 Heating/Cooling/Humidification Malfunctions

결함 증상 가능한 원인들 해법
난방 없음 1. Over-temperature protection tripped; 2. Solid-state relay failure; 3. Heating element failure 1. Reset the over-temperature protection; 2. Use a multimeter to measure the output terminals of the solid-state relay (replace if there is no voltage); 3. Measure the resistance of the heating element (several tens of ohms at room temperature; replace if the resistance is infinite).
Heating Runaway (Continuous Temperature Rise) 1. Stuck solid-state relay; 2. Sensor short circuit; 3. Controller malfunction 1. Immediately turn off the power and replace the solid-state relay; 2. Inspect the sensor (replace it if there is a short circuit); 3. Contact customer service to have the controller repaired.
Not cooling 1. The cooling switch is not turned on; 2. The compressor has not started; 3. There is a refrigerant leak 1. Go to “Operation Settings” and turn on the cooling switch; 2. Check the compressor power supply and phase sequence, then wait 3 to 5 minutes before restarting; 3. Contact customer service to locate the leak and recharge the refrigerant
느린 냉각 1. Low refrigerant level; 2. Clogged condenser; 3. Excess moisture in the chamber 1. Top off the refrigerant (to the amount specified for the equipment); 2. Clean dust from the condenser; 3. Dry the moisture from the chamber before lowering the temperature
No humidification 1. Low water level in the tank; 2. Humidifier malfunction; 3. Water level controller failure 1. Check the tap water supply and ensure the water level in the tank meets the required standard; 2. Measure the resistance of the humidifier (replace if abnormal); 3. Replace the water level controller

7.4 Other Common Problems

결함 증상 가능한 원인들 해법
The case is live 1. Poor grounding; 2. Live wire touching the chassis; 3. Electrical leakage from the heater 1. Re-ground the unit and ensure the resistance is ≤50 Ω; 2. Inspect the internal wiring and secure the live wire terminals; 3. Measure the heater’s insulation resistance; replace it if it is defective.
Poor door seal (temperature and humidity leakage) 1. The weatherstripping has deteriorated; 2. The cabinet door hinges are loose; 3. The door lock is improperly adjusted 1. Replace the weatherstripping; 2. Adjust the hinge bolts to ensure the door is vertical; 3. Turn the door lock knob to tighten the cabinet door
The circulation fan is not running 1. Fan power supply failure; 2. Motor failure 1. Check the fan wiring to ensure it is powered; 2. Use a multimeter to test the motor windings (if there is no open circuit, replace the motor)

Appendix: Circuit Diagram

High And Low Temperature Humidity Chamber-Figure24 High And Low Temperature Humidity Chamber-Figure25 High And Low Temperature Humidity Chamber-Figure26 High And Low Temperature Humidity Chamber-Figure27 High And Low Temperature Humidity Chamber-Figure28

부록 및 다운로드

비디오

GDJS-015B 고온 및 저온 습도 환경 챔버 - 작동 비디오