고온 압력 시험 장치
사용자 설명서
적용 할 수있는 LISUN models: GNGPL-3610、GNGPL-3610-2PA、GNGPL-3610-3PA
1. 장비에 대한 기본 이해
1.1 기술 사양
| LISUN 모델 | GNGPL-3610 | GNGPL-3610-2PA |
| 워크스테이션 수 | 1 | 2 |
| 잎 | 1 round blade (Ф6) | One rectangular blade and one Ф6 circular blade |
| 블레이드 두께 | 0.7 mm | |
| 무게 | 2.5N weight (including the weight of the blade): 1 | 2.5N weights (including the weight of the blade): 2 |
| 적용 기준 | BS 1363-1 그림 10 | GB/T 2099.1, 조항 30.1 및 그림 41(IEC 60884-1, Clause 30.1 and Figure 44) |
Note: This device must be used in conjunction with the LISUN GW-225 High Temperature Chamber and is specifically designed for high-temperature pressure testing of insulation sheaths on plugs and prongs for household and similar purposes.
1.2 응용 시나리오
Plug and Socket Manufacturers: Inspection of incoming plastic materials, 100% final inspection of finished products, and comparative testing of new materials; Third-Party Safety Testing Laboratories: CCC certification, CE (IEC) certification, and mandatory testing for UK BS certification; Quality Inspection / Market Regulatory Authorities: Random inspections of plug products in e-commerce and brick-and-mortar markets; Small Appliance Manufacturers: In-house reliability verification of plugs for adapters, rice cookers, electric kettles, and desk lamps.
1.3 Download Electronic Documents
전자 사용자 설명서, 교정 인증서, 보증서 및 기타 문서의 다운로드 링크는 제품 발송 시 이메일로 이미 보내드렸습니다. 가능한 한 빨리 다운로드해 주시기 바랍니다. 만약 이메일을 받지 못하셨다면, 저희에게 연락하여 다운로드 링크를 다시 요청해 주십시오.
2. 안전상의 주의
During high-temperature testing, temperatures may reach up to 200°C. When opening the door to remove or place test specimens, wear heat-resistant gloves; do not touch the inner chamber, blades, weights, or fixtures with bare hands to prevent burns.
Weights are heavy metal objects; handle them gently. Do not throw or stack them. The blades are sharp; be careful of the cutting edges when installing or replacing test specimens to avoid cutting your hands. When moving the entire press fixture, lift it steadily to prevent the blades from causing scratches.
3. 장비개요
3.1 Standard Diagram
그림 3-1을 참조하십시오(다음에서 발췌). IEC 60884-1, Figure 44).

그림 3-1
3.2 Photographs of the Actual Items
3.2.1 GNGPL-3610
그림 3-2를 참조하세요.
| 번호 | 주의 사항 |
| ① | 라운드 블레이드 |
| ② | 무게 |
| ③ | 조정 가능한 기초 |

그림 3-2
3.2.2 GNGPL-3610-2PA
그림 3-3를 참조하세요.
| 번호 | 주의 사항 |
| ① | Rectangular Blade |
| ② | 라운드 블레이드 |
| ③④ | 무게 |
| ⑤ | 조정 가능한 기초 |

그림 3-3
4. High-Temperature Pressure Test of the Insulation Sheath
4.1 Preparation of Specimens and Apparatus
Rectangular blades are used for round-pin tests; circular blades are used for other pin tests. Refer to Section 3.1 to position the specimen correctly.
4.2 시험 절차
Set the high-temperature chamber to 200°C and allow it to stabilize at the target temperature. Place the test setup, along with the test specimen mounted in its normal position, in the high-temperature chamber for 2 hours.
Next, remove the test apparatus and the specimen, detach the specimen from the apparatus, and immerse it in cold water within 10 seconds to cool it.
4.3 시험 결과 평가
Measure the thickness of the insulating material at the indentation point; the reduction in thickness compared to the original measurement at the start of the test should not exceed 50 percent.
4.4 Post-Test Procedures
4.4.1 장치 재설정
After the test is complete, remove the weights and blades from the test rig bracket. Do not allow the weights and blades to remain suspended on the test rig for extended periods when the test is not in progress.
4.4.2 샘플 준비
시험에 합격한 샘플과 불합격한 샘플을 분리하고, 명확하게 표시하여 혼동을 피하십시오.
5. 장비의 일상 유지보수 및 관리
5.1 Daily Maintenance After Shutdown
At the end of each day’s testing, clean dust and sample debris from the equipment surfaces, and keep the chucks and slide rails clean and free of debris; check that the blades are not deformed.
5.2 Maintenance During Long-Term Storage
After routine maintenance is complete, apply a neutral rust-preventive oil to the test apparatus, blades, weights, and other components to prevent rust; cover the entire unit with a dust cover and store it in a dry, well-ventilated warehouse.
5.3 교정 주기 요구사항
측정 정확도와 장비의 작동 신뢰성을 보장하기 위해, 자격을 갖춘 제3자 계측 연구소에 정기적으로 보내 교정을 받는 것이 좋습니다. 권장 교정 주기는 12개월입니다. 사용자는 사용 빈도, 환경 조건 및 품질 시스템 요구 사항을 고려하여 구체적인 교정 주기를 결정할 수 있습니다.
6. 문제 해결 및 일반적인 문제 해결 방법
| 결함 증상 | 가능한 원인들 | 순서 |
| The blade is not perpendicular to the sample | The test setup is not level. | Please adjust the base of the test apparatus so that it is level. |

中文简体