스프링 임팩트 해머 교정 장치

사용자 설명서

적용 할 수있는 LISUN 모델: IK01-07-CAL

1. 장비에 대한 기본 이해

1.1 주요 매개변수

1.1.1 Specifications Table

파라미터 주의 사항
Calibrating the Impact Element 250 g (pre-installed), 500 g (spare)
교정 범위 0~2J
Repeatability ±0.01 J
Swinging the arm can cause damage <0.002 J

The equipment consists of three components: the main frame, the energy indication system, and the release positioning system. The overall structure complies with the requirements of Figure B.1 in Appendix B of GB/T 2423.55-2023 (IEC 60068-2-75:2014).

1.1.2 Structural Diagram

Equipment configuration, as shown in Figure B.1 of Appendix B of GB/T 2423.55-2023 (IEC 60068-2-75:2014). See Figure 1-1.

Spring Impact Hammer Calibrating Device-Figure1

그림 1-1

1.2 장비 구성 검증

1.2.1 포장 풀기

When unpacking the equipment, handle it gently to avoid scratching the exterior with sharp tools. After unpacking, first inspect the exterior of the main unit to ensure there are no obvious deformations, dents, or loose parts. If you notice any abnormalities with the equipment or accessories, please contact us immediately.

A piano wire with a pointer is installed at the location marked in the figure below. Please exercise special care when removing the packaging to avoid damaging it.

Spring Impact Hammer Calibrating Device-Figure2

그림 1~2

1.2.2 본체 및 부속품

그림 1~3을 참조하십시오.

번호 주의 사항
The IK01-07-CAL Spring Impact Hammer Calibrating Device can be used to directly calibrate 250g impact elements.
Used to calibrate the 500g impact element; for backup purposes; not required when calibrating the 0–2 J spring hammer.

Spring Impact Hammer Calibrating Device-Figure3

그림 1~3

2. 안전상의 주의

2.1 Pre-Operation Safety Requirements

Operators must be familiar with the entire contents of this manual; unauthorized operation without proper training is prohibited;

The equipment must be placed on a stable, level, and vibration-free surface. Use a spirit level to ensure it is level, and avoid tilting or wobbling;

It is strictly prohibited to disassemble the core components of the device (release mechanism, pendulum, and hands), and unauthorized adjustments to its structure are prohibited.

2.2 Safety Requirements During Operation

The spring-loaded impact hammer must be triggered via the equipment’s release mechanism; direct manual release of the impact hammer is strictly prohibited to prevent injury and damage to the equipment;

When operating the machine, keep your hands away from the impact zone and the area where the arm swings to avoid being pinched or struck;

After completing a single calibration, wait until the pendulum arm has come to a complete stop, reset the pointer to zero, and then proceed to the next operation.

3. 장비개요

그림 3-1를 참조하세요.

번호 주의 사항
① Arm swing c Absorbs the impact of the hammer and transfers the energy to the hand
② Bearing a
③ Pointer b Indicates the impact energy value as the pendulum arm deflects
④ Energy Gauge f According to the text prompts on the energy dial, refer to the numbers on the outer ring to calibrate the 250-gram impact element; the unit is J.
⑤ Release base d Standard impact hammer with automatic trigger; manual release is prohibited.
⑥ Impact contact point k Impact hammer strike points, ensuring consistent force distribution
⑦ Spring-loaded hammer slot Secure the calibrated spring-loaded impact hammer to ensure precise positioning.
⑧ Release mechanism (turntable) e Push the spring-loaded hammer forward until the spring-loaded hammer is released and the base automatically releases.
⑨ Spirit Level Before using the equipment, make sure it is level and securely positioned.
⑩ Footed glass Four adjustable feet allow the device to be leveled

Spring Impact Hammer Calibrating Device-Figure4

그림 3-1

4. Calibration Test for the Spring-Loaded Impact Hammer

4.1 Installation and Alignment of the Impact Hammer

As shown in Figure 4-1, place the spring-loaded impact hammer—with the spring fully coiled—into the slot, ensuring it is securely fastened and does not wobble;

Rotate the turntable to align the impact hammer with the impact contact point k, ensuring precise positioning.

Spring Impact Hammer Calibrating Device-Figure5

그림 4-1

4.2 Single Impact and Reading

Continue turning the dial to advance the impact hammer; the release mechanism on the base will automatically trigger the release of the impact arm. Do not release it manually;

As shown in Figure 4-2, after the pendulum arm swings to its maximum angle and returns, the pointer remains at its maximum deflection position. Read the value on the outer ring of the energy scale indicated by the pointer and record it as the first data point;

Spring Impact Hammer Calibrating Device-Figure6

그림 4-2

Manually reset the pointer to zero, allow the swing arm to hang naturally, and reset the turntable before performing the next operation.

4.3 Three Standard Procedures

After completing the first test, rotate the impact hammer to a new position, repeat steps 4.1–4.2, and record the data from the second test;

Change the position of the impact element again, repeat the procedure, and record the data from the third test;

For all three operations, ensure that the installation, triggering, and reading conditions are consistent;

4.4 Data Recording and Calculations

Actual impact energy (J) = (1st reading + 2nd reading + 3rd reading) ÷ 3. Compare the calculated average to the impact hammer’s rated energy to determine whether the test passes.

5. 장비의 일상 유지보수 및 관리

5.1 일상 청소 및 유지 관리

After each use, wipe down the device’s surface, energy dial, card slot, and release base with a clean, soft cloth to remove dust and grease;

Do not use corrosive solvents such as acetone to wipe the energy scale or the surface of the pointer;

Lubricate the rotating parts—such as bearings, swing arms, and the release mechanism (turntable)—with a small amount of special-purpose lubricant every 6 months to ensure smooth operation.

5.2 Periodic Inspection of Core Components

Pointers: Check monthly for deformation or looseness; if they cannot be reset, calibrate them promptly;

Release Mechanism (Turntable): Check monthly to ensure the trigger operates smoothly, with no jamming or looseness;

Swing Arm and Bearings: No deformation, no abnormal wear, and no resistance during rotation.

5.3 Long-Term Storage and Protection

When the equipment is not in use for an extended period, cover it with a dust cover and store it in a dry, well-ventilated, vibration-free environment;

Do not stack heavy objects on top of each other; do not store them together with sharp or hard objects;

Avoid direct sunlight to prevent the energy gauge from fading and components from deteriorating.

5.4 교정 주기 요구사항

측정 정확도와 장비의 작동 신뢰성을 보장하기 위해, 자격을 갖춘 제3자 계측 연구소에 정기적으로 보내 교정을 받는 것이 좋습니다. 권장 교정 주기는 12개월입니다. 사용자는 사용 빈도, 환경 조건 및 품질 시스템 요구 사항을 고려하여 구체적인 교정 주기를 결정할 수 있습니다.

6. 문제 해결 및 일반적인 문제 해결 방법

결함 증상 가능한 원인들 순서
읽기 오류 The equipment is not level; the bearings are seized. Refer to Section 2.1 to place the equipment on a level surface; refer to Section 5.1 to ensure that all components move smoothly. If any components are stuck, they must be repaired or recalibrated.