전기 케이싱 압축 시험 장치
사용자 설명서
적용 할 수있는 LISUN 모델: LS-B21
1. 장비에 대한 기본 이해
1.1 주요 매개변수
1.1.1 장비 기술 사양
| 파라미터 | 주의 사항 |
| 제어 방법 | Touchscreen + PLC Control |
| 압축 방식 | 전기 |
| Compression Force Value | 50–1000 N, adjustable |
| 센서 정확도 | FS 1 % |
| Compress Time | Adjustable from 1 to 99 minutes (seconds); once the compression time has elapsed, the system automatically releases pressure and retracts. |
| 변위 해상도 | 1 mm |
1.1.2 적용 가능한 표준
The LS-B21 Electrical Casing Compression Test Device strictly complies with GB/T 17466.1-2019″Electrical Accessories—Installation Boxes and Enclosures for Fixed Electrical Installations for Household and Similar Purposes—Part 1: General Requirements,” specifically Clauses 15.3 and 15.5 (IEC 60670-1:2021 Clauses 15.2 and 15.5), and is used to conduct testing.
It is primarily used in electrical appliance quality inspection laboratories, quality control departments of manufacturers, third-party testing organizations, and for import and export product inspections.
1.2 장비 구성 검증
1.2.1 포장 풀기
기기를 개봉할 때는 날카로운 도구로 외부 케이스에 흠집이 생기지 않도록 조심스럽게 다루십시오. 개봉 후에는 먼저 기기 외관을 검사하여 눈에 띄는 변형이나 찌그러짐이 없는지, 버튼이 손상되었거나 헐거워지지 않았는지 확인하십시오. 기기 또는 부속품에 이상이 발견되면 전원을 켜지 마시고 즉시 저희에게 연락해 주십시오.
1.2.2 본체 및 부속품
| 번호 | 주의 사항 |
| ① | LS-B21 Compression Test Device Main Unit |
| ② | 전원 코드 |
1.2.3 전자 문서 다운로드
We have already sent download links for the electronic user manual, calibration certificate, warranty card, and other documents via email at the time of shipment. Please download them promptly. If you missed the email, you may also contact us directly to request the download links again.
2. 안전상의 주의
2.1 전기 안전
Before connecting the power cord, verify that the supply voltage is AC 220 V ± 10% and the frequency is 50 Hz or 60 Hz;
You must use a three-prong outlet with a protective ground, with a protective ground resistance of ≤4 Ω. Under no circumstances should you cut or disconnect the protective ground wire, as doing so poses a risk of electric shock.
2.2 환경 안전
Do not use in environments with flammable or explosive substances, strong electromagnetic interference, high temperatures and humidity, or excessive dust levels;
Operating Environment: Temperature 5°C to 35°C, relative humidity ≤ 65% RH.
2.3 운영 절차
Improper operation can easily result in personal injury, equipment damage, and distorted test data; operators must thoroughly read this manual and strictly follow the procedures;
Remove jewelry before performing the procedure, tie back long hair, and do not wear loose-fitting clothing;
During the test, take precautions to prevent test specimen fragments from flying and causing injury;
During the test, do not extend your hands or tools into the fixture’s travel range to prevent crush injuries or accidents;
Operators must not leave their posts without authorization while the equipment is in operation; in the event of an abnormality, immediately shut off the power supply;
Do not perform tests that exceed the load capacity or stroke limit, and do not use damaged or non-standard fixtures.
2.4 기타
본 설명서에 나와 있지 않은 사항에 대해서는 주의를 기울여 진행하시거나 당사에 문의해 주십시오.
3. 장비개요
그림 3-1를 참조하세요.
| 번호 | 주의 사항 |
| ① | Power Cord Connection and Power Switch |
| ② | 터치 스크린 |
| ③ | Specimen Mounting Position |

그림 3-1
4. 터치스크린 프로그램 소개
4.1 초기화면
그림 4-1를 참조하세요.
| 번호 | 주의 사항 |
| ① | Display current force, displacement, deformation, and maximum force |
| ② | If necessary, enter the sample information. |
| ③ | Test Plan Setup |
| ④ | Display Settings: You can set the units for force values. |
| ⑤ | Curve Display Area |
| ⑥ | Up/Down buttons: Manually adjust the size of the specimen mounting position based on the sample size. |
| ⑦ | Click to clear the current test data |
| ⑧ | Click to return the ram to its initial position |
| ⑨ | Stop Test Button, Start Test Button |

그림 4-1
4.2 Test Plan Interface
그림 4-2를 참조하세요.
| 번호 | 주의 사항 |
| ① | To test the speed, you can use the default value. |
| ② | Break-induced shutdown: The machine automatically shuts down when the break reaches a set percentage. Optional feature |
| ③ | Test Type: The standard accessories included with this unit are intended for compression testing only. Please select “Compression.” |
| ④ | Start deformation calculation; calculation begins once the actual force reaches a specified percentage of the set force value. Optional setting |
| ⑤ | Shutdown conditions, optional pressure values. Optional settings |
| ⑥ | Start the crush test; once the actual force reaches a certain percentage of the set force, begin calculating the fracture ratio. Optional setting |
| ⑦ | 강제 설정 |

그림 4-2
4.3 Force Value Settings Screen
그림 4-3를 참조하세요.
| 번호 | 주의 사항 |
| ① | Set the required force value |
| ② | Set the pressure duration |
| ③ | Speed reduction settings: When the pressure reaches a set percentage, the speed slows down. Use the default values. |
| ④ | Pressure-activated timing start setting: Timing begins when the actual pressure reaches 100% of the set pressure value. |

그림 4-3
5. Compression Test Procedure
5.1 Sample Preparation and Placement
Select a specimen of an electrical enclosure that is free of initial cracks, plastic deformation, and in perfect condition. Based on the specimen dimensions and in accordance with Section 4.1, use the up/down keys to adjust the platen to a position slightly higher than the specimen’s height, and place the specimen in the specimen mounting position as described in Chapter 3.
5.2 테스트 매개변수 설정
Set the compression force and duration in accordance with Section 4.3 and the relevant provisions of the standards.
Clause 15.3 of GB/T 17466.1-2019 (IEC 60670-1:2021 Clause 15.2) requires a force of 500 N applied for 1 minute; GB/T 17466.1-2019, Clause 15.5 (IEC 60670-1:2021, Clause 15.5) requires a force of 50 N/cm² (not exceeding 1000 N) to be applied through 9 mm thick plywood, with a duration of 1 minute.
5.3 Starting the Compression Test
| 번호 | 주의 사항 |
| ① | Click the “Start Test” button to begin the test |
| ② | Move the press block slowly toward the specimen and apply pressure. |
| ③ | When the actual pressure reaches the set value, the timer starts. |
| ④ | Once the force application time reaches the set duration, the test stops automatically and the load block returns to its original position. |
Note: If the actual pressure does not reach the set value, or if the specimen breaks during the compression process, you may click “Stop Test” to manually terminate the test.
5.4 시험 결과 해석
After the test is completed, the mounting box and housing must not exhibit any deformation or damage that would affect continued use; if this is the case, the test is considered passed.
6. 장비의 일상 유지보수 및 관리
6.1 일상적인 유지보수
After each test, clean up any dust or debris from the specimen mounting area to keep the work surface clean and free of debris; wipe down the equipment surface to keep it clean; verify that the clamping block is secure and that the limit stops are not misaligned; apply specialized lubricant to the drive screw to ensure smooth operation.
6.2 Storage Specifications
6.2.1 저장 환경 요구 사항
Temperature: 5°C to 35°C; relative humidity: ≤65% RH; dry, dust-free environment free of corrosive gases.
6.2.2 장기 저장 작업
전원을 끄고 모든 케이블을 분리하십시오. 장비를 청소하고 방진 필름으로 감싸십시오. 매달 30분씩 전원을 켜서 부품의 습기 손상을 방지하십시오.
6.3 교정 주기 요구사항
To ensure the measurement accuracy and operational reliability of the equipment, it is recommended that it be sent periodically to a qualified third-party metrology laboratory for calibration. The recommended calibration interval is 12 months. The specific calibration interval may be determined by the user based on frequency of use, environmental conditions, and quality system requirements.
7. 문제 해결 및 일반적인 문제 해결 방법
| 결함 증상 | 가능한 원인들 | 순서 |
| The fixture does not raise or lower | Limit Switch Triggered | Check and adjust the limit switches |
| Force and displacement values are displayed incorrectly | Not reset to zero | Click to reset and try again |
| The machine stops as soon as the specimen is subjected to a load | Parameter Configuration Error | See Section 4.2; do not set unnecessary parameters arbitrarily. |