+8618117273997웨이 신
영어영어
中文简体 中文简体 en English ru Русский es Español pt Português tr Türkçe ar العربية de Deutsch pl Polski it Italiano fr Français ko 한국어 th ไทย vi Tiếng Việt ja 日本語

직류 바이어스 전류원을 사용하는 LCR 미터

사용자 설명서

적용 할 수있는 LISUN 모델: LS1373CX、LS1373X,LS1379CMX,LS1379MX

The LS1373 is an LCR Meter with a built-in high-frequency LCR digital bridge and a DC bias current source ranging from -12 to 12 A. With a current resolution as low as 0.125 mA, it can analyze the characteristics of inductor cores for inductor quality inspection or product characteristic analysis. The built-in bias current source features lower noise and more flexible control, and includes an integrated comparator, making it easy to integrate into automated production lines for inductor testing.

1. 제품 개요

This chapter introduces the product positioning and key performance characteristics of the LS1373 LCR Meter with DC Bias Current Source.

1.1 성능 특징

Provides a maximum constant current output of 12A per unit

0.125 mA fine-tuned current increments

Forward/Reverse Switching Function for Superimposed Current

Supports automatic current continuation after the circuit is opened

Supports the ability to set the duration of the current boost

Two current output modes: constant current and step scanning

Graphical user interface, available in both Chinese and English

Built-in comparator, making it easy to integrate into automated production lines for inductor testing

2. 기술 사양

2.1 Technical Specifications of the Superimposed Current Source

기술 사양 기술 사양 기술 사양 기술 사양 기술 사양
모델 모델 LS1373 LS1373 LS1373
7 ″ LCD 7 ″ LCD 7 ″ LCD
행정부 행정부 Physical buttons + foot switch Physical buttons + foot switch Physical buttons + foot switch
Supported Measurement Frequencies Supported Measurement Frequencies 50 Hz–500 kHz; C-type: custom-made up to 1 MHz 50 Hz–500 kHz; C-type: custom-made up to 1 MHz 50 Hz–500 kHz; C-type: custom-made up to 1 MHz
전류 범위 전류 범위 -12A ~ 12A -12A ~ 12A -12A ~ 12A
현재 정확도 현재 정확도 ±1% ±5 mA ±1% ±5 mA ±1% ±5 mA
Current 측정 범위 0mA~0.500A 0.500A ~ 2.500A 2.5 A–12 A
Current 단계별 0.125 mA 0.625 mA 3.125 mA
최대 출력 전압 최대 출력 전압 11V (Outputs 11V under all current conditions) 11V (outputs 11V under all current conditions) 11V (Outputs 11V under all current conditions)
Maximum Allowable DC Resistance Maximum Allowable DC Resistance Rmax = Vmax/I (Ω) Rmax = Vmax/I (Ω) Rmax = Vmax / I (Ω)
Maximum Allowable Inductance Maximum Allowable Inductance Lmax = Vmax / (di/dt) (mH) Lmax = Vmax / (di/dt) (mH) Lmax = Vmax / (di/dt) (mH)
Range Selection Method Range Selection Method Automatic Automatic Automatic
Start and Stop Control Methods Start and Stop Control Methods START/STOP physical buttons, 4 types of foot controls, bus START/STOP physical buttons, 4 types of foot controls, bus START/STOP physical buttons, 4 types of foot controls, bus
Maximum Continuous Current Load Duration Maximum Continuous Current Load Duration Runs for 2 to 3 hours with continuous output Runs for 2 to 3 hours with continuous output Runs for 2 to 3 hours with continuous output
인터페이스 인터페이스 RS232C, USB Host, USB Device, Handler RS232C, USB Host, USB Device, Handler RS232C, USB Host, USB Device, Handler

2.2 Technical Specifications of the LCR Meter

기술 사양 기술 사양 기술 사양 기술 사양 기술 사양 기술 사양
모델 모델 LS1373 LS1373A LS1373B LS1373C
테스트 빈도 테스트 빈도 50Hz ~ 200kHz 50Hz ~ 300kHz 50Hz ~ 500kHz 50Hz ~ 1MHz
테스트 빈도 테스트 빈도 More than 12000 frequency points More than 12000 frequency points More than 12000 frequency points More than 12000 frequency points
테스트 매개 변수 테스트 매개 변수 C, L, X, B, R, G, D, Q, ESR, Rp, θ, |Z|, |Y| C, L, X, B, R, G, D, Q, ESR, Rp, θ, |Z|, |Y| C, L, X, B, R, G, D, Q, ESR, Rp, θ, |Z|, |Y| C, L, X, B, R, G, D, Q, ESR, Rp, θ, |Z|, |Y|
테스트 레벨 테스트 레벨 50 mV to 2 V rms 50 mV to 2 V rms 50 mV to 2 V rms 50 mV to 2 V rms
정확성 정확성 0.1% (LCR connection method only; basic accuracy at 1 kHz, at the midpoint of the measurement range) 0.1% (LCR connection method only; basic accuracy at 1 kHz reference range) 0.1% (LCR connection method only; basic accuracy at 1 kHz reference range) 0.1% (LCR connection method only; basic accuracy at 1 kHz, near full scale)
속도 테스트 속도 테스트 Fast: 50 times/second, Medium: 20 times/second, Slow: 6 times/second (at frequencies above 1 kHz without current applied) Fast: 50 times/second, Medium: 20 times/second, Slow: 6 times/second (at frequencies above 1 kHz without current) Fast: 50 times/second, Medium: 20 times/second, Slow: 6 times/second (at frequencies above 1 kHz without current applied) Fast: 50 times/second, Medium: 20 times/second, Slow: 6 times/second (at frequencies above 1 kHz without current applied)
표시 범위 |Z|, R, X 0.01 mΩ to 99.9999 MΩ 0.01 mΩ to 99.9999 MΩ 0.01 mΩ to 99.9999 MΩ 0.01 mΩ to 99.9999 MΩ
표시 범위 |Y|, 지, 비 0.00001 μS to 99.9999 S 0.00001 μS to 99.9999 S 0.00001 μS to 99.9999 S 0.00001 μS to 99.9999 S
표시 범위 C 0.00001 pF to 9.99999 F 0.00001 pF to 9.99999 F 0.00001 pF to 9.99999 F 0.00001 pF to 9.99999 F
표시 범위 D 0.00001-9.99999 0.00001-9.99999 0.00001-9.99999 0.00001-9.99999
표시 범위 L 0.00001 μH to 99999.9 H 0.00001 μH to 99999.9 H 0.00001 μH to 99999.9 H 0.00001 μH to 99999.9 H
표시 범위 Q 0.01-99999.9 0.01-99999.9 0.01-99999.9 0.01-99999.9
표시 범위 θ (RAD) 3.14159까지 3.14159 3.14159까지 3.14159 3.14159까지 3.14159 3.14159까지 3.14159
표시 범위 θ(도) -179.999 ° ~ 179.999 ° -179.999 ° ~ 179.999 ° -179.999 ° ~ 179.999 ° -179.999 ° ~ 179.999 °
표시 범위 Δ% -99.999 % ~ 99.999 % -99.999 % ~ 99.999 % -99.999 % ~ 99.999 % -99.999 % ~ 99.999 %
신호 소스 출력 임피던스 신호 소스 출력 임피던스 30Ω, 100Ω, 1062X 30Ω, 100Ω, 1062X 30Ω, 100Ω, 1062X 30Ω, 100Ω, 1062X
Range Selection Method Range Selection Method Auto, Hold Auto, Hold Auto, Hold Auto, Hold
트리거 방법 트리거 방법 Internal, Manual, Automatic, External Internal, Manual, Automatic, External Internal, Manual, Automatic, External Internal, Manual, Automatic, External
교정 기능 교정 기능 Open Circuit, Short Circuit, Full-Frequency Reset, Spot-Frequency Reset Open Circuit, Short Circuit, Full-Frequency Reset, Spot-Frequency Reset Open Circuit, Short Circuit, Full-Frequency Reset, Spot-Frequency Reset Open Circuit, Short Circuit, Full-Frequency Reset, Spot-Frequency Reset
등가 회로 등가 회로 직렬, 병렬 직렬, 병렬 직렬, 병렬 직렬, 병렬
비교기 비교기 Three-tier sorting: BIN 0–BIN 2 Three-tier sorting: BIN 0–BIN 2 Three-tier sorting: BIN 0–BIN 2 Three-tier sorting: BIN 0–BIN 2
비교기 비교기 PASS, HI, LOW (PASS/FAIL LED indicator) PASS, HI, LOW (PASS/FAIL LED display) PASS, HI, LOW (PASS/FAIL LED indicator) PASS, HI, LOW (PASS/FAIL LED display)

3. 이용약관

3.1 전원 공급 장치

LCR Meter With DC Bias Current Source-Figure1

Factory default power supply voltage: 100–242 VAC;

Power Supply Frequency: 46–64 Hz

Power consumption: ≤300 VA

3.2 주변 온도 및 습도

Normal operating temperature: 0°C to 40°C; humidity: < 90% RH

Reference operating temperature: 20°C ± 8°C; humidity: < 80% RH

Transportation ambient temperature: 0°C to 55°C; humidity: 93% RH

3.3 예열

After turning it on, let it warm up briefly before use.

3.4 주의사항

LCR Meter With DC Bias Current Source-Figure2

(1) Please do not use this product in adverse environments such as those with dust, vibration, direct sunlight, or corrosive gases.

(2) If the instrument will not be used for an extended period, please place it in its original packaging or a similar box and store it in a well-ventilated room at a temperature between 5°C and 40°C and a relative humidity of no more than 85% RH. The air should be free of harmful impurities that could corrode the instrument, and direct sunlight should be avoided.

(3) This instrument has been carefully designed to minimize noise interference from the AC input power supply; however, it should still be used in a low-noise environment whenever possible. If this is unavoidable, please install a power filter.

3.5 부피와 무게

Width × Height × Depth: 395 mm × 135 mm × 465 mm

무게 : 9.9의 kg

3.6 안전 요구 사항

This instrument is a Class I safety instrument.

(1) Insulation Resistance

Under reference operating conditions, the insulation resistance between the power supply terminals and the enclosure shall be no less than 50 MΩ;

Under humid and hot transport conditions, the insulation resistance between the power terminals and the enclosure shall be no less than 2 MΩ;

(2) Dielectric Strength

Under reference operating conditions, the power supply terminals and the enclosure can withstand an AC voltage of 1.5 kV at a frequency of 50 Hz for 1 minute without breakdown or arcing.

(3) Leakage current

The leakage current must not exceed 3.5 mA.

3.7 전자기 호환성

(1) Power supply transient immunity shall comply with the requirements of GB 6833.4.

(2) Conduction sensitivity shall comply with the requirements of GB 6833.6.

(3) Radiated interference shall comply with the requirements of GB 6833.10.

4. Panel Description

4.1 Front Panel Description

 

LCR Meter With DC Bias Current Source-Figure3

Figure 2-1 Front Panel

1. Trademarks and Model Numbers

Instrument brand, model number, measurement range, etc.

2. LCD 디스플레이

Extra-large 800×480 dot-matrix LCD display that shows settings, measurement results, measurement conditions, and more.

3. Soft Keys

Six soft keys are available for selecting controls and parameters; the function of each soft key is defined to the left of it. The soft key definitions vary depending on the display page.

4. PASS LED

The test results are indicated by an LED light.

5. NEXT button

The fast page-turn keys are only useful during the Voltage Transformer scan test function.

6. FAIL LED

The test failed due to the LED light.

7. Number Keys

Used to enter numbers and other relevant information.

8. The ESC key

Finish entering the numbers on the numeric keypad.

9. [←] key

The BACKSPACE key. Press this key to delete the last character of the entered value.

10. KEYLOCK Key

Press the [KEYLOCK] button; the [KEYLOCK] indicator will light up, indicating that the panel buttons are currently locked. Press the [KEYLOCK] button again, and the [KEYLOCK] button will turn off, indicating that the keypad lock has been released. If the password function is set to “ON,” you must enter the correct password to release the keypad lock; otherwise, the keypad lock cannot be released.

When the instrument is controlled via a bus (such as USBTMC), the [KEYLOCK] button will light up. Press the [KEYLOCK] button again, and it will turn off, indicating that the instrument has returned to local mode and the keyboard lock has been released.

11. [DC BIAS] Button

Used to enable or disable the DC bias output. Press the [DC BIAS] key; the [DC BIAS] indicator will light up, indicating that the DC bias output is enabled; Press the [DC BIAS] key again, and the [DC BIAS] key will turn off, indicating that DC bias output is disabled. On certain non-test screens where DC bias cannot be applied, pressing this key will have no effect.

12. [RESET] button

Press the [RESET] button to stop the scan while the Voltage Transformer is being automatically scanned; the instrument will not perform any operations on other pages.

13. [TRIGGER] key

When the instrument’s trigger mode is set to MAN (Manual) mode, press this key to trigger the instrument to perform a test.

14. [ENTER] key

When entering values, press the [ENTER] key to complete data entry, confirm, and save the data displayed on the bottom line of the LCD.

When entering a filename, press the [ENTER] key to complete the entry, confirm, and save the filename displayed on the bottom line of the LCD.

15. Chassis Ground Terminal

This terminal is connected to the instrument housing. It can be used for protective or shielding ground connections.

16. DC bias current output terminal

Used to connect to the inductor under test and apply a DC bias current to it

DC Bias Current, Positive Terminal (DRIVE+)

DC Bias Current (Negative Terminal) DRIVE-

17. Test End (UNKNOWN)

Four-terminal to test terminal. Used to connect a four-terminal test fixture or test cable for measuring the device under test.

High-side current drive (Hcur);

High-side voltage sampling (Hpot);

Low-end voltage sampling (Lpot);

Low-end current drive (Lcur).

18. [SYSTEM] Menu Key

Press the [SYSTEM] button to access the “System Settings” page.

19. [SETUP] Menu Button

Press the [SETUP] button to enter the “Measurement Settings” main screen.

20. [LCRZ] Menu Key

Press the [LCRZ] key to enter the “Component Measurement Display” page.

21. [RETURN] Menu Key

Back button: On certain settings pages, press this button to return to the previous menu.

22. Power Switch (POWER)

Press this button to turn the power on or off.

23. USB HOST Port

Used to connect a USB flash drive.

4.2 Rear Panel Description

 

LCR Meter With DC Bias Current Source-Figure4

Figure 2-2: Rear Panel

1. EXT. TRG

Foot switch connector.

2. HANDLER Port

The HANDLER interface makes it easy to set up automated test systems and perform automated testing, and is widely used in automated factory production lines. The instrument uses this interface to output range comparison results and status signals, and also receives the “Start” signal through this interface.

3. USBDEVICE Interface

The USBDEVICE interface enables a computer to control the LCR digital bridge.

When the bus mode is set to USBTMC, this port functions as a USBTMC interface.

When the bus mode is set to USBCDC, this port functions as a USBCDC interface (virtual serial port).

When the bus mode is set to USBHID, this port functions as a USBHID interface (virtual keyboard interface).

4. RS232C 직렬 인터페이스

Serial communication interface, which allows the computer to control the instrument.

5. 환기창

Heat dissipation to maintain the instrument at its normal operating temperature.

6. Electrical Outlet

Used to connect to an AC power source.

7. Ground Terminal

This terminal is connected to the instrument housing. It can be used for protective or shielding ground connections.

8. Barcode Labeling Area

Used for affixing instrument barcodes.

4.3 Screen Areas: Names and Functions of Each Section

The LS1373 features a 7-inch LCD display with a resolution of 800×480, and the functional layout of its display area is shown in Figure 2-3.

 

LCR Meter With DC Bias Current Source-Figure5

Figure 2-3 shows the functional zoning of the area

Display Page Area

This area displays the name of the current page.

Document Field

Move the cursor to this area to perform file management operations. File management operations include: loading, saving, and deleting.

도구 섹션

Some less commonly used features, for which there are no corresponding settings fields on the display page, are listed in the “Tools” section.

Soft Key Area

This area is used to display the function definitions for soft keys. The function definitions for soft keys vary depending on the position of the cursor within the field.

Function Display Area

This area displays the test conditions and functions that have been set.

측정 결과 영역

This area displays test result information

Message Prompts and Data Entry Areas

This area is used to display system prompts and user data entries.

사용에 대한 지침

To help users get started quickly, this user manual provides operating instructions only for the LS1373’s superimposed current test function. For operating instructions regarding the LCR bridge functions, users should refer to our company’s *LCR Digital Bridge General User Manual*.

5. 작동 지침

5.1 Installation, Connection, and Testing

5.1.1 Connecting the Test Cable

5.1.1.1 Connect the inductor under test using the LS1373Y03 test cable

 

LCR Meter With DC Bias Current Source-Figure6

This test cable combines the DC and AC current terminals into a single unit, which is soldered to the terminals of a 4-terminal clip for user convenience. However, because it is designed to handle high currents, this cable does not use shielded wire and is therefore not suitable for high frequencies. It is recommended that the test frequency not exceed 20 kHz.

The connection method for performing an open-circuit reset using the LS1373Y03 test cable is shown in the figure below:

 

LCR Meter With DC Bias Current Source-Figure7

Generally, open-circuit zeroing is not required; however, when the impedance |Z| of the inductor under test exceeds 1 kΩ and the test requires high accuracy, open-circuit zeroing is necessary.

The connection method for performing a short-circuit reset using the LS1373Y03 test cable is shown in the figure below:

 

LCR Meter With DC Bias Current Source-Figure8

Generally, short-circuit zeroing is not necessary; however, when the impedance |Z| of the inductor under test is less than 10 Ω, short-circuit zeroing is required for tests with high accuracy requirements.

Note: When measurement stability of 10 nH is required, the short-circuit effect achieved by clamping the test probes together is not as effective as clamping them onto the two terminals of a gold-plated shorting pad that are positioned very close together.

5.1.1.2 Testing the Inductor Under Test Using a Cable Connection with LS28Y11 + LS1373Y02

 

LCR Meter With DC Bias Current Source-Figure9

For this test connection method, use the LS1373Y02 test lead for applying current and the LS28Y11 LCR test cable separately; short-circuit both leads at the end of the inductor under test. Although this wiring method is slightly more complicated, it can be used for high-frequency testing. When the test frequency exceeds 20 kHz, this wiring configuration is recommended.

The connection method for performing an open-circuit reset using the LS28Y11+LS1373Y02 test cable is shown in the figure below:

 

LCR Meter With DC Bias Current Source-Figure10

Generally, open-circuit zeroing is not necessary; however, when the impedance |Z| of the inductor under test exceeds 1 kΩ and the test requires a high degree of accuracy, open-circuit zeroing must be performed.

The connection method for short-circuit reset using the LS28Y11+LS1373Y02 test cable is shown in the figure below:

 

LCR Meter With DC Bias Current Source-Figure11

Generally, short-circuit zeroing is not necessary; however, when the impedance |Z| of the inductor under test is less than 10 Ω, short-circuit zeroing is required for tests with high accuracy requirements.

Note: When measurement stability of 10 nH is required, the short-circuit effect achieved by clamping the test probes together is not as effective as securing the test probes in a fixed position where the two clamps are very close to each other on the gold-plated shorting pad.

5.1.1.3 Testing Non-Inductive Components Using the LCR Bridge Function of the LS1373

When users wish to test non-inductive components, such as resistors or capacitors, using the LCR bridge function of the LS1373 alone, they must connect the resistor or capacitor under test to the LS28Y11 separately. They should not simultaneously apply DC bias current through the DRIVE+ and DRIVE- terminals, to prevent the superimposed current from affecting measurement accuracy.

Please refer to the figure below:

 

LCR Meter With DC Bias Current Source-Figure12

★ Note: When using the bridge function to test resistors and capacitors, do not connect the DRIVE+ and DRIVE- terminals for the DC bias current.

5.2 Operating the Main Functions

5.2.1 Bias Current

 

LCR Meter With DC Bias Current Source-Figure13

In the “Component Measurement Display” page’s bias section, users can set the desired DC bias current.

The specific parameters for the DC bias current of the LS1373 아래 표에 나와 있습니다.

전류 범위 전류 범위 -12A ~ 12A -12A ~ 12A -12A ~ 12A
현재 정확도 현재 정확도 ±1% ±5 mA ±1% ±5 mA ±1% ±5 mA
Current 측정 범위 0mA ~ 0.5A 0.5A ~ 2.5A 2.5 A–12 A
Current 단계별 0.125 mA 0.625 mA 3.125 mA

5.2.2 Starting the Measurement with the Bias Current

After setting the bias current and other required test conditions, and connecting the device under test according to the test lead connection method described in the previous section, press the [DC BIAS] button on the front panel. The [DC BIAS] indicator will light up, indicating that the DC bias current is active.

 

LCR Meter With DC Bias Current Source-Figure14

Once the measurement is complete, turn off the DC bias by pressing the [DC BIAS] button. The [DC BIAS] indicator will turn off, and the DC bias current will be turned off.

Offset Current Hold Time

Some inductors can overheat and burn out when subjected to high currents. The LS1373 can prevent this by controlling the current application time through the bias current hold time setting. For details, see the bias current hold time settings.

자동 갱신 기능

On the production floor, to improve testing efficiency, it is not desirable to turn the current on and off after testing each inductor. The LS1373 allows you to replace one inductor with another after the current terminal is open-circuited without having to press a button to turn the power off and then back on; the instrument automatically completes this process, making it convenient to use. This feature must be enabled on the < System Configuration> page.

5.2.3 내부 저항

 

LCR Meter With DC Bias Current Source-Figure15

In the “Internal Resistance” section of the < Measurement Settings> page, users can set the source internal resistance for the desired LCR function.

The inductance of many inductors depends on frequency and on the magnitude of the AC current flowing through the inductor during measurement. Once the measurement frequency and level are fixed, the value of the inductance is determined by the source impedance. To facilitate data comparison with instruments from other manufacturers on the market, the LS1373 offers three internal resistance modes, as detailed in the table below:

Internal Resistance Mode Primary Comparison Instruments
30 Ω *8511D*
100 Ω *4284A, *4980A
1062X * 1062 *

The instrument is set to 1062X mode by default; this mode provides better stability for superimposed current testing.

However, the 1062X mode supports frequencies up to a maximum of 300 kHz; for frequencies above 300 kHz, we recommend using the 30 Ω internal resistance mode.

5.2.4 Offset Current List Scan Settings

 

LCR Meter With DC Bias Current Source-Figure16

On the < List Scan Settings> page, users can select and configure DC bias current scanning.

5.2.5 Offset Current List Scan Display

 

LCR Meter With DC Bias Current Source-Figure17

On the < List Scan Settings> page, after the user selects and configures a DC bias current scan, they proceed to the < List Scan Display> page. When the [DC BIAS] button is pressed to apply the current, the instrument automatically performs a sequential current scan and measurement according to the current settings previously configured by the user.

5.2.6 Automatic Flow Resumption Feature:

 

LCR Meter With DC Bias Current Source-Figure18

In the “Automatic Flow Resumption” section of the < Test Configuration> page, users can enable or disable the instrument’s automatic flow resumption feature.

Feature Description: On the production floor, to improve testing efficiency, it is undesirable to have to turn the current on and off each time a new inductor is tested. The LS1373 automatically handles this process—when the current terminal is open-circuited and another inductor is inserted—without requiring the user to press a button to turn the current off and then back on, making it convenient to use.

떨어져서

When “Auto Renew” is set to OFF, on the measurement display page, with DC BIAS enabled, if the instrument’s current drive terminal is open-circuited, the instrument will automatically cut off the current output relay and disable DC BIAS.

ON

When the Auto Hold setting is set to ON, on the measurement display screen, with DC BIAS enabled, if the instrument’s current drive terminal is open-circuited, the instrument will not automatically deactivate the current output relay or turn off DC BIAS. Instead, it will automatically adjust the drive current to a value close to 0 A. When the user reconnects the inductor under test, the instrument will automatically adjust the output current to the user-set value.

In addition, when the test clamps are open, the instrument displays “OVER LOAD,” indicating that the current source is overloaded and the measurement cannot proceed normally.

5.2.7 Setting the Offset Current Hold Time:

 

LCR Meter With DC Bias Current Source-Figure19

In the “Current Hold Time” section of the < Test Configuration> page, users can set the current hold time. If set to “0,” the instrument assumes this feature is not in use.

Function Description: Some inductors can overheat and burn out when a high current is applied. The LS1373 can control the duration of current application by setting the bias current hold time, thereby preventing such issues.

5.2.8 Open-Circuit and Short-Circuit Reset Function:

 

LCR Meter With DC Bias Current Source-Figure20

When the impedance |Z| of the inductor under test exceeds 1 kΩ, or when the test requires a high degree of accuracy, an open-circuit zeroing procedure is necessary.

The method for opening the circuit varies depending on the type of fixture; please refer to the descriptions in the previous sections.

When the impedance |Z| of the inductor under test is less than 10 Ω, or when the test requires a high degree of accuracy, a short-circuit zeroing procedure must be performed.

The short-circuiting method varies depending on the type of fixture; please refer to the descriptions in the previous sections.

LS1373 recommends using [Open-Circuit Single-Frequency Clear] and [Open-Circuit Single-Frequency Clear] to clear open-circuit and short-circuit conditions.

5.2.9 Common Steps for Measuring Current Through an Inductor:

1. 전원 켜기

2. Set the measurement conditions

3. Perform open-circuit and short-circuit calibration based on the impedance of the inductor under test

4. Connect the inductor under test to the test clamps.

5. Turn on the DC bias output

6. View the measurement results on the measurement page

7. Turn off the DC bias output.

5.3 Command Reference

This instrument allows you to control the current level, set the startup current, and check the instrument’s status using SCPI commands via the RS232C interface. The relevant commands are as follows:

5.3.1: BIAS: CURRENT

Syntax: BIAS: CURRent < numeric>

: BIAS: CURRent ?

This setting configures the DC bias current. Setting this value does not mean that the DC bias is enabled. Because this series of instruments offers a high degree of flexibility, the instrument does not strictly verify whether the current value set by the user is valid. Please note this.

파라미터

  < Numeric>
범위 12까지 12
단위 A

5.3.2: BIAS: STATe

Syntax: BIAS: STATE {ON|OFF|1|0}

: BIAS: STATe?

Describes the activation of DC bias. DC bias is automatically turned off after the state is loaded from memory. When DC bias is set to ON, the output provides the DC bias set by : BIAS: CURRent.

파라미터

  기술설명
ON or 1 Enable DC Offset
OFF or 0 Disable DC Offset

5.3.3: BIAS: ERROR

Syntax: BIAS: ERROR?

Describes the current error status of the query bias source.

파라미터

  기술설명
0 No errors occurred
1 OVERLOAD or UNBAL
2 과부하
3 UNBAL

6. HANDLER Interface User Guide

The instrument provides users with a Handler interface, which makes it easy to build an automated component sorting system. When the instrument is used in an automated component sorting and testing system, this interface provides signals for interacting with the system and for outputting sorting results.

6.1 Definition of the HANDLER Interface

Output Signal Specifications: Low-level active, open-collector output, optically isolated.

 

LCR Meter With DC Bias Current Source-Figure21

Handler Interface Definition

The flowchart for Bin Sorting when sorting into three bins is shown in the figure:

Definition of 3-Level Sorting Signals

발 위치 기능
1 /NG (D/QNG) is a non-pass signal; the output is active low.
2 /P1 is a first-class product that meets specifications; the output is low and valid.
3 /P2 indicates a second-grade合格 product with low output effectiveness
4 /P3 is a Grade 3合格 product with low output effectiveness
5 WAIT (Busy) output, active-high. This signal can be used to control the operation of external mechanical processing equipment during instrument testing; when this signal is active, the instrument is performing measurements and calculations.
6 /EOC (A/D End): Output, active low. When this signal is output, the measurement (A/D conversion) of the current device under test has been completed, although calculations may still be in progress. When this output is high, good contact between the test terminals and the current device under test must be ensured; when it is low, external mechanical equipment is permitted to move the next test device to the test terminals in preparation for the next measurement.
7 /START (Start): Input; valid only if it is a falling edge and the low-level duration is greater than 2.5 ms. This signal is supplied externally to the HANDLER input.
8 EXTV External Input Power Supply (The instrument is connected to the external power source by default)
9 Low-side output of the COM interface

Signal Mapping Chart for the DB9 Socket Connector of the Handler Interface

The instrument’s operating sequence is shown in the figure below:

 

LCR Meter With DC Bias Current Source-Figure22

Instrument Timing Diagram

T1: The duration of the START signal must be greater than 2.5 ms. Instrument measurement begins when the STARTU signal transitions from high to low, and the signal must be removed before T2 returns to low; otherwise, another measurement may be triggered erroneously. If no external START signal is present, pressing the start button on the instrument’s front panel will also generate a start signal.

T2: The instrument’s measurement and calculation time; at the end of T2, the sorting results are sent to the HANDLER interface.

T3: The wait time from when WAIT is inactive to when START is active. Determined by the mechanical transmission mechanism.

T4: Instrument A/D conversion time.

As shown in the figure, the valid interval for outputting the measurement and sorting results for each component is the period from the end of the current T2 to the end of the next T2.

6.2 Wiring Between the Instrument and the PLC

Generally speaking, the necessary drive circuits are already built into both the instrument and the PLC; all the user needs to do is connect the 24V, 0V, and any required signal lines.

7. 포장 목록

     
1 LS1373 주인 1
2 Spare fuse (2A) 2
3 LS28Y11 Test Line 1
4 LS1373Y03 Test Line 1
5 전원 코드 1
6 CD-ROM (containing the electronic user manual, warranty card, certificates, etc.) 1