+8618117273997웨이 신
영어영어
中文简体 中文简体 en English ru Русский es Español pt Português tr Türkçe ar العربية de Deutsch pl Polski it Italiano fr Français ko 한국어 th ไทย vi Tiếng Việt ja 日本語

고니오포토미터용 감쇠기

사용자 설명서

적용 할 수있는 LISUN 모델: LSG-1800A,LSG-1800ACCD,LSG-1890B,LSG-1890BCCD

1. 제품 개요

1.1 사용 사례

This attenuator is a dedicated accessory for the LSG series of Goniophotomoeters. It is used to reduce excessively strong light signals, bringing them within the probe’s normal measurement range, thereby ensuring that testing can proceed normally for high-power luminaires with narrow beam angles or in scenarios with short test distances.

During a normal test, the software displays the following prompt (Figure 1-1), indicating that an attenuator must be used when a signal overflows; otherwise, the test cannot be completed.

Attenuator For Goniophotometer-Figure1

그림 1-1

1.2 주요 기능

Neutral optical attenuation reduces only the light intensity without altering other parameters of the light signal; the product is factory-calibrated to ensure precise attenuation coefficients; it can be used individually or in combination with multiple units; compatible with LSG series Goniophotomoeter probes for easy installation.

2. 제품 소개

2.1 Introduction to Attenuators

그림 2-1를 참조하세요.

번호 주의 사항
Attenuator Serial Number
Attenuation Coefficient of an Attenuator

Attenuator For Goniophotometer-Figure2

그림 2-1

Note: Each attenuator has a unique serial number and attenuation coefficient; please take this into account when using them.

2.2 Rules of Use

When used individually, the attenuation coefficient can be applied directly; when used in combination, the total attenuation coefficient equals the product of the individual coefficients. Example: Z212055T × Z212045T = 7.373 × 7.223 = 53.255179

Note: The software only accepts three decimal places; the fourth decimal place is rounded. For example, if the coefficient is 53.255179, enter 53.255.

2.3 Principles of Use

Do not use attenuators unless necessary: Do not install them when the signal is normal to avoid unnecessary attenuation; do not use multiple attenuators unless necessary: Use a single attenuator whenever possible to minimize cumulative error.

3. Installation and Use

3.1 Installation of Attenuators

Remove the light shield from the front of the probe, rotate the attenuator clockwise to attach it to the probe connector, and ensure it is securely fastened, as shown in Figures 3-1 and 3-2.

Attenuator For Goniophotometer-Figure3

그림 3-1

Attenuator For Goniophotometer-Figure4

그림 3-2

If you need to install multiple attenuators, stack the second and subsequent attenuators clockwise in the same manner. Once all attenuators are installed, reattach the light shield, as shown in Figure 3-3.

Attenuator For Goniophotometer-Figure5

그림 3-3

3.2 Setting the Attenuation Coefficient

Press the PM400C device settings button, as shown in Figure 3-4.

Attenuator For Goniophotometer-Figure6

그림 3-4

Click “Next Page” on the screen, as shown in Figure 3-5.

Attenuator For Goniophotometer-Figure7

그림 3-5

Refer to Figure 3-6 and locate the “Filter Coefficient” option. When no attenuator is used, the default coefficient is 1.000. When an attenuator is used, enter the corresponding precise coefficient. Each time you add or remove an attenuator, you must promptly adjust the attenuation coefficient.

Attenuator For Goniophotometer-Figure8

그림 3-6

3.3 테스팅

Refer to the Goniophotomoeter user manual and perform the test as usual.

4. 정기 유지보수 및 관리

4.1 Cleaning Optical Surfaces

Gently wipe the optical surface only with lens paper or a lint-free cloth; do not touch it with paper towels or your fingers. If there are stains, dip a small amount of anhydrous alcohol and wipe gently; do not scratch or bump the optical surface.

4.2 보관 및 취급

When not in use, store it in its original protective case in a dry, dust-free environment; avoid high temperatures, humidity, and corrosive gases; it is recommended to include a desiccant; protect it from being crushed or dropped to prevent damage to the optical surface.

4.3 정기 검사

Before use, inspect the optical surface for scratches, cracks, or mold; if any are found, stop using it immediately and replace it with a new attenuator.

4.4 교정 주기 요구사항

측정 정확도와 장비의 작동 신뢰성을 보장하기 위해, 자격을 갖춘 제3자 계측 기관에 정기적으로 보내 교정을 받는 것이 좋습니다. 권장 교정 주기는 12개월입니다. 사용자는 사용 빈도, 환경 조건 및 품질 시스템 요구 사항을 고려하여 구체적인 교정 주기를 결정할 수 있습니다.

5. Troubleshooting and Resolution

결함 증상 가능한 원인들 순서
A “signal overflow” error still appears after installing the attenuator. PM400C Coefficient Error, Insufficient Attenuation Coefficient Refer to Section 3.2 to confirm that the coefficient has been corrected; install an attenuation plate to increase the attenuation coefficient.
The test data showed significant deviations after using the attenuator. PM400C Coefficient Error, Excessive Use of Attenuators, Damaged Attenuators Refer to Section 3.2 to verify that the attenuation coefficient has been corrected; reduce the number of attenuators to lower the attenuation coefficient while ensuring there is no signal overflow; check the attenuators for dirt or damage, and replace them if necessary.

부록 및 다운로드

비디오

LSG-1800A 고정밀 회전 조명기구 Goniophotometer - 작동 비디오

LSG-1890B 고정밀 회전 조명기구 Goniophotometer - 작동 비디오

LSG-1800B 고정밀 회전 조명 기구 Goniophotometer - 작동 비디오