고정밀 회전 조명기구 Goniophotometer

사용자 설명서

적용 할 수있는 LISUN 모델: LSG-1890B

1. 제품 개봉 및 기본 사용법 익히기

1.1 Unboxing and Inspection

When unpacking the device, handle it gently to avoid scratching the casing with sharp objects. After unboxing, first inspect the exterior of the device to ensure there is no visible deformation or dents, and that the buttons are not damaged or loose. If you notice any abnormalities with the device or its accessories, do not turn it on. Please contact us immediately.

1.2 시스템 구성 및 개요

1.2.1 핵심 하드웨어 구성 요소

참고: 아래 표는 전체 시스템의 핵심 하드웨어 구성 요소를 나열합니다. 구매하신 장비에는 이러한 구성 요소 중 일부만 포함될 수 있으므로, 이 설명서의 지침을 따르실 때 이 점을 유의하시기 바랍니다.

구성 요소 이름 모델 핵심 기능
고니오포토미터 메인 유닛 LSG-1890B Used in conjunction with a photometric probe to measure photometric parameters and their spatial distribution, such as luminous flux and light distribution curves.
근접장 광도 측정 프로브 PM400C(N) Suitable for standard darkrooms, such as those around 10 meters in size
Far-Field Photometric Probe PM400C(화씨) 선택 사항입니다. 길이가 20미터가 넘는 매우 긴 암실에 적합합니다.
표준 조명 SLS-150W Used to calibrate the darkroom (luminance sensor)
E27 Fixture / C-감마 테스트용; 백열전구에 적합
크로스 클램프 For C-Gamma testing; suitable for other indoor and outdoor lighting fixtures
사각 클램프 B-베타 테스트용; 투광등에 적합
크로스 라인 레이저 Used during installation and commissioning; used during calibration and testing to determine the position and orientation of the standard lamp and the sample luminaire
내각 CASE-19 Used to house angle controllers, power meters, and DC and AC power supplies. Includes a power strip, an RS485 communication module, and more.
각도 조절기 RT-200C The angle at which the Goniophotomoeter is positioned for display and control. Generally, no manual adjustment is required; control is handled by the software.
파워 미터 LS 시리즈 시료의 전압, 전류, 전력 및 역률과 같은 전기적 매개변수를 측정합니다.
DC 전원 공급 장치 DC 시리즈 직류 샘플, 표준 램프 및 보조 램프에 안정적인 직류 전원을 공급합니다.
AC 전원 LSP Series AC 샘플에 안정적이고 순수한 정현파 AC 전원을 공급합니다.

1.2.2 Software, manuals, certificates, etc.

At the time of shipment, a download link containing all the documents was sent via email. Please download them promptly. If you missed the email, you can contact us to request a new download link.

소프트웨어, 사용자 설명서, 보증 카드, 표준 램프 교정 인증서 및 기타 중요 문서와 관련 문서가 포함됩니다.

1.4 측정 매개변수

Illuminance data, illuminance distribution, area luminous flux, luminaire efficiency, luminance distribution (optional), utilization factor, luminance limit curve, glare rating, isophote curves, maximum allowable height-to-distance ratio, luminaire curve vs. illumination area, isointensity curves, effective beam angle, EEI, UGR, PAR, PPF, and PPFD spatial distribution tests, etc.

2. 중요 사항

2.1 전기 안전

The system must be connected to a stable AC power source rated at 220 V ±5% with a frequency of 50 Hz or 60 Hz. Ensure proper grounding with a grounding resistance of ≤4 Ω. Do not use the equipment if the power connections are loose or the cables are damaged.

2.2 환경 안전

광학 암실을 사용할 때는 주변 온도를 25±1°C로 유지해야 하며, 상대 습도는 10%~65% RH 사이로 유지해야 합니다.

제어실과 암실 모두 건조하고 환기가 잘 되며 먼지가 없고 부식성 가스가 없도록 유지해야 합니다.

Avoid placing the device near strong magnetic or electric fields; it must be placed on a level, sturdy surface.

2.3 Darkroom Operating Procedures

2.3.1 암실 설명

Please confirm that the darkroom has been constructed exactly according to the darkroom drawings we provided. If there are any discrepancies in the actual dimensions, please inform us before installing and commissioning the equipment so that we can adjust the plan, ensuring that the installation and commissioning proceed smoothly and that the test results are accurate;

The interior of the darkroom must be completely painted matte black—that is, a non-reflective pure black—including all walls, ceilings, floors, partitions, doors, and so on. A small number of components, such as lighting fixtures, air conditioning units, and power switches, may not be painted black, provided they do not appear in the optical path between the Goniophotomoeter’s main unit and the photometric probe;

Once the equipment has been installed and commissioned, do not move any of the hardware components inside the darkroom, including the main unit base, the photometer probe base, the photometer probe orientation, and the cross-laser base. Doing so may result in inaccurate test results.

2.3.2 암실 사용 시 주의사항

암실에 들어갈 때는 먼지 유입을 방지하기 위해 신발을 벗거나 신발 덮개를 착용하십시오.

The sample luminaire must be securely mounted to prevent it from falling and damaging the Goniophotomoeter during testing;

Do not place any objects on the main unit of the Goniophotomoeter, and do not place any objects within the Goniophotomoeter’s rotation range;

During testing, the darkroom must remain sealed to prevent external light from interfering with the test results.

2.3.3 Precautions When the Darkroom Is Not in Use

암실은 항상 닫아 두십시오. 허가받지 않은 사람은 출입할 수 없습니다. 먼지가 암실에 들어가지 않도록 하십시오.

2.4 Standard Lamp Operating Procedures

표준 램프는 수명이 1년이며 설치 및 시운전 중 암실에서 광도계 프로브를 교정하는 데 사용됩니다.

실험실에서 암실을 매년 교정해야 하는 경우, 새 표준 램프를 구입할 수 있습니다.

스탠드형 램프는 사용 직후 뜨거우니 만지지 마십시오. 램프 케이스에 다시 넣기 전에 10분 정도 식히십시오.

The standard lamp used in the darkroom is a directional standard lamp, and its direction is controlled by its bracket. Do not remove the bulb from its bracket; doing so will render the standard lamp inoperable.

2.5 기타

본 설명서에 나와 있지 않은 사항에 대해서는 주의를 기울여 진행하시거나 당사에 문의해 주십시오.

3. Hardware Assembly

3.1 캐비닛 장비 조립

3.1.1 Introduction to the Cabinet

캐비닛을 평평하고 견고한 작업대 위에 놓고 작업에 필요한 충분한 공간을 확보하십시오.

번호 주의 사항
배선도: 배선도에 따라 장비를 조립하십시오.
배전반은 캐비닛 내부의 모든 기기 전원 코드를 연결하는 데 사용됩니다.
냉각 팬 전원 코드: 캐비닛 내부의 멀티탭에 연결해 주세요.
RS-485 Communication Box, used to connect all devices via an RS-485 communication cable
Please connect the power adapter for the RS485 communication box to the power strip inside the cabinet.
The USB cable for the RS485 communication box must be connected to the computer.
RT-200C 각도 조절기 배선
LS Series Power Meter Wiring
DC 전원 연결
AC 전원 배선
The main power cable for the cabinet, which supplies power to all equipment inside the cabinet, should be routed out through cable opening ② and connected to the power source only after all equipment has been fully assembled.
Cable opening—the main power cord can be routed through here. Note: The main power cord plug is a Chinese standard plug; you may need to provide your own power adapter.

High Precision Rotation Luminaire Goniophotometer-Figure1

그림 1

3.1.2 RT-200C Installation

For now, all you need to do is plug it in.

High Precision Rotation Luminaire Goniophotometer-Figure2

그림 2

3.1.3 LS 전력계 설치

번호 주의 사항
Connected via an RS-485 communication cable; the other end is connected to the RS-485 communication box inside the cabinet.
전원 코드가 연결되어 있습니다. 다른 한쪽 끝은 이미 캐비닛 내부의 배전반에 연결되어 있습니다.
V1 test lead, connected to the red terminal of the power meter’s voltage V
V2 test lead, connected to the black terminal of the power meter’s voltage V
A1 테스트 리드는 전력계의 전류 A 단자(빨간색 단자)에 연결됩니다.
A2 테스트 리드는 전력계의 전류 A용 검은색 단자에 연결됩니다.

High Precision Rotation Luminaire Goniophotometer-Figure3

그림 3

3.1.4 DC 전원 공급 장치 설치

번호 주의 사항
Connected via an RS-485 communication cable; the other end is already connected to the RS-485 communication box inside the cabinet.
직류 전원 출력선을 연결할 때 양극과 음극 단자가 일치하는지 확인하십시오.
전원 코드가 연결되어 있습니다. 다른 한쪽 끝은 이미 캐비닛 내부의 멀티탭에 연결되어 있습니다.

High Precision Rotation Luminaire Goniophotometer-Figure4

그림 4

3.1.5 교류 전원 설치

Please locate the following AC power output cable in the accessories:

High Precision Rotation Luminaire Goniophotometer-Figure5

그림 5

번호 주의 사항
Connected via an RS485 communication cable; the other end is already connected to the RS485 communication box inside the cabinet.
전원 코드가 연결되어 있습니다. 다른 한쪽 끝은 이미 캐비닛 내부의 멀티탭에 연결되어 있습니다.
Run the AC power output cable through cable opening ④ and connect it to position ⑤.
와이어 구멍
AC power output cable: Pass it through cable hole ④ and connect it to position ③.

High Precision Rotation Luminaire Goniophotometer-Figure6

그림 6

High Precision Rotation Luminaire Goniophotometer-Figure7

그림 7

3.2 Installation of The Goniophotomoeter Main Unit and Goniophotomoeter Probe

3.2.1 호스트 조립

The same number is engraved on the main unit’s base tapered sleeve ① and the corresponding position on the main unit ②. Rotate the tapered sleeve until the two numbers are aligned, then secure the tapered sleeve using the hex socket screws provided as standard.

High Precision Rotation Luminaire Goniophotometer-Figure8

그림 8

High Precision Rotation Luminaire Goniophotometer-Figure9

그림 9

Remove the screws securing the dust cover on the turntable, and pull the dust cover off.

High Precision Rotation Luminaire Goniophotometer-Figure10

그림 10

Use a forklift to lift the turntable; since the turntable is quite heavy, three to four people can also carefully lift it. Align the dust cover on the turntable with the terminal block on the base, then slowly lower the turntable onto the base. Hold the terminal block in place with your hand so that it slides into the turntable along with the tapered sleeve, preventing the terminal block from being crushed by the turntable.

High Precision Rotation Luminaire Goniophotometer-Figure11

그림 11

턴테이블의 방향을 미세 조정하려면 아래 나열된 순서대로 단계를 따르십시오.

번호 주의 사항
턴테이블의 나사 구멍 6개와 받침대의 나사 구멍 6개를 맞춰 정렬한 후, 제공된 육각 나사와 와셔를 사용하여 턴테이블을 받침대에 고정하십시오.
참고: 모든 표준 와셔를 반드시 사용하십시오. 그렇지 않으면 나사가 제대로 맞지 않아 턴테이블 회전 중 장비가 손상될 수 있습니다.
Plug the base terminal block into the corresponding terminal block socket on the turntable.
Then put the dust cover back in place

High Precision Rotation Luminaire Goniophotometer-Figure12

그림 12

High Precision Rotation Luminaire Goniophotometer-Figure13

그림 13

본체 하단에 있는 네 개의 다리는 높이 조절이 가능합니다.

번호 주의 사항
Turn the screw at the bottom to raise the leg until it is off the ground.
동시에 맨 위쪽 나사가 다른 곳에 눌리지 않도록 주의하세요. 눌리지 않는다면 살짝 풀어도 됩니다.

High Precision Rotation Luminaire Goniophotometer-Figure14

그림 14

Position the goniophotomoeter in the correct location according to the darkroom blueprints. Align the four sides of the base as closely as possible with the darkroom walls. Then remove the auxiliary arm.

High Precision Rotation Luminaire Goniophotometer-Figure15

그림 15

Next, please turn all four legs of the Goniophotomoeter until they are in contact with the floor.

3.2.2 광도 측정 프로브 및 프로브 홀더 조립

예를 들어, 광도계가 하나밖에 없는 암실을 생각해 보세요.

Assemble the photometric probe and probe mount, and position them in the appropriate locations according to the darkroom drawings; there is no need to secure them with screws for now.

High Precision Rotation Luminaire Goniophotometer-Figure16

그림 16

High Precision Rotation Luminaire Goniophotometer-Figure17

그림 17

3.2.3 시스템 배선

번호 주의 사항
The four terminal blocks ①on the cabinet are connected to the Goniophotomoeter⑩; simply match them one-to-one based on the numbers.
The two terminals connect to the terminal block inside the cabinet and are used to supply power to all equipment in the darkroom, including the Goniophotomoeter and photometric probes. This allows for convenient control of all darkroom equipment via the cabinet’s power switch.
참고: 암실 장비는 전원 콘센트에서 직접 전원을 공급받을 수도 있습니다.
캐비닛 접지 단자: 전선을 사용하여 접지에 연결하십시오.
캐비닛 전원 코드는 220V, 50/60Hz 전원에 연결됩니다.
고니오포토미터 본체용 전원 코드
고니오포토미터의 접지 단자를 전선을 사용하여 접지에 연결하십시오.
This is a one-to-two communication cable; one end of the RS-485 cable connects to the RS-485 communication box inside the cabinet, and the other end connects to position ⑭ on the back of the RT-200C.
광도계 프로브 트리거 와이어는 프로브 ⑬에 연결하십시오. 프로브가 하나만 설치된 경우 기본 설정은 근거리(N)입니다.
If you have only one photometer in your darkroom, there is no need to connect the far-field (F) sensor.
The four terminal blocks ⑩on the Goniophotomoeter are connected to the cabinet①; simply match them one-to-one according to the numbers.
Power Adapter for Photometric Probe
광도계 프로브용 RS-485 통신 케이블이며, 다른 한쪽 끝은 캐비닛 내부의 RS-485 통신 박스에 연결됩니다.
광도계 프로브 트리거 케이블은 고니오광도계의 본체 ⑧에 연결됩니다.
RT-200C 통신 케이블의 다른 한쪽 끝은 고니오포토미터의 메인⑦ 장치에 연결됩니다.

High Precision Rotation Luminaire Goniophotometer-Figure18

그림 18

High Precision Rotation Luminaire Goniophotometer-Figure19

그림 19

High Precision Rotation Luminaire Goniophotometer-Figure20

그림 20

High Precision Rotation Luminaire Goniophotometer-Figure21

그림 21

4. 소프트웨어 설치 및 구성

4.1 소프트웨어 및 드라이버 설치

4.1.1 소프트웨어 런타임 환경

System Requirements: Windows 7/8/10/11 (32-bit/64-bit); the computer must have at least one USB port.

4.1.2 설치

번호 주의 사항
LSG 시리즈 고니오포토미터 소프트웨어 - 두 번 클릭하여 설치하세요
This APK file can be copied to an Android phone for installation. The app is intended solely for rotating the Goniophotomoeter and cannot be used for testing. Since the software includes all the app’s features, this manual uses the software as an example for operation.
RS485 Communication Box Driver—Please double-click to install

High Precision Rotation Luminaire Goniophotometer-Figure22

그림 27

4.1.3 예외 처리

If your antivirus software or the Windows Firewall flags a file as suspicious during the software installation process, please temporarily disable the antivirus software or the Windows Firewall and try installing the software again.

If the software fails to open when you double-click its icon after a successful installation, please check whether you are logged in to your computer with an administrator account. If the software still does not open after logging in with an administrator account, right-click the software icon to open its Properties window. On the Compatibility tab, check the box next to “Run this program as an administrator,” click “Apply,” and then double-click the icon again to open the software.

High Precision Rotation Luminaire Goniophotometer-Figure23

그림 28

4.2 소프트웨어 설정

First, turn on the cabinet’s power switch, then power on all system devices, including all equipment inside the cabinet, the Goniophotomoeter main unit, and the photometric probes. Double-click to launch the software. The first time you run the software, it will automatically open the system configuration interface. Select the correct model for each device you have purchased.

번호 주의 사항
고니오포토미터 모델을 선택하려면 오른쪽에 있는 드롭다운 목록을 클릭하십시오. 구매하신 모델은 다음과 같습니다. LSG-1890B.
근거리 광도계 프로브; 광도계 프로브 모델 PM400C를 선택하십시오.
Far-field photometric probe—if you have not purchased one, please select “Not Installed.”
The LSG-1890B does not include a near-field Spectroradiometer; please select “Not Installed.”
AC power supply models are generally from the LSP Series
DC power supply models are generally designated as the DC Series
Power meter model: The model number is printed on the front panel of the power meter. Please check it and select the appropriate model.
Once you have selected all the correct models, click “Auto-detect Port.”
After the search for all ports is complete, click “OK” to save and exit.

High Precision Rotation Luminaire Goniophotometer-Figure24

그림 29

Note: If a communication error occurs during the automatic port search, please read the error message carefully, check whether the corresponding device is powered on, and verify that the RS485 communication cable is properly connected. Once you have confirmed these points, click “Automatic Port Search” again until communication is established with all devices.

After clicking OK, the software automatically opens the database file save dialog. From then on, all test reports will be automatically saved in this database file.

번호 주의 사항
데이터베이스 파일을 저장할 경로를 선택하십시오. 바탕화면이나 바탕화면 폴더에는 저장하지 마십시오.
파일 이름을 입력하세요
"저장"을 클릭하세요.

High Precision Rotation Luminaire Goniophotometer-Figure25

그림 30

5. Calibration of darkroom equipment (Each step must be performed carefully and thoroughly; otherwise, calibration data may be inaccurate and test results may be abnormal.)

5.1 주 장치의 수평 시운전

5.1.1 레이저 준비 완료

고니오포토미터의 메인 암 뒷면에 레이저 스위치가 있습니다. 라인 레이저 스위치를 켜면 메인 암 바로 앞쪽으로 붉은색 라인 레이저 빔이 방출됩니다.

High Precision Rotation Luminaire Goniophotometer-Figure26

그림 31

십자선 레이저 장치와 삼각대를 찾으세요.

High Precision Rotation Luminaire Goniophotometer-Figure27

그림 32

Remove the cross-line laser and the batteries. Insert the batteries into the cross-line laser. The power switch is located on the side of the device. The device turns on when the switch is set to the “unlocked” position.

번호 주의 사항
전원 스위치: 전원을 켜려면 스위치를 "잠금 해제" 위치로 돌리십시오.
VH는 각각 수직 및 수평 레이저 스위치를 의미합니다.

High Precision Rotation Luminaire Goniophotometer-Figure28

그림 33

십자선 레이저를 삼각대에 장착하십시오.

Note: The cross-line laser is self-leveling. If the cross-line laser is not self-leveled, an audible alarm will sound. This may be due to an uneven surface or a tripod that is not properly set up. Please ensure that the cross-line laser remains self-leveled at all times during use.

5.1.2 Steps for Horizontal Commissioning of the Host Machine

암실 도면에 따르면, 중심 높이는 다음과 같습니다. LSG-1890B 고니오포토미터의 높이는 1510mm입니다. 삼각대를 사용하여 십자선 레이저에서 나오는 수평 레이저 빔의 높이를 1510mm로 조정하십시오. 이 높이에서 수평 레이저 빔은 암실 칸막이에 있는 모든 광투과 구멍의 중앙을 통과해야 합니다.

High Precision Rotation Luminaire Goniophotometer-Figure29

그림 34

주의: 회전 중 장비 손상을 방지하기 위해 고니오포토미터 위 또는 회전 범위 내에 어떠한 물체도 올려놓지 마십시오.

소프트웨어를 엽니 다.

번호 주의 사항
클릭하여 각도 모드를 전환하세요
①을 클릭하면 B-베타 각도 대신 C-감마 각도가 표시됩니다.
그런 다음 클릭하여 각도 제어 인터페이스를 엽니다.

High Precision Rotation Luminaire Goniophotometer-Figure30

그림 35

Click the continuous rotation button for any direction. The Gamma axis of the Goniophotomoeter will then begin to rotate continuously.

High Precision Rotation Luminaire Goniophotometer-Figure31

그림 36

At this point, observe the horizontal laser beams emitted by the single-line and cross-line lasers from the main axis of the Goniophotomoeter.

High Precision Rotation Luminaire Goniophotometer-Figure32

그림 37

Then, using an appropriate wrench, raise the goniophotomoeter and level it by adjusting the height of its four legs. The final result of the calibration: Regardless of the Gamma angle of the goniophotomoeter, its straight laser beam perfectly aligns with the horizontal laser beam, as shown in the figure below.

High Precision Rotation Luminaire Goniophotometer-Figure33

그림 38

You’ll need plenty of patience to complete this step. Once adjusted, make sure all four legs of the main unit are firmly on the ground. Then click the “Stop” button on the software’s angle control interface to stop the Goniophotomoeter from rotating. Next, lock the four legs of the Goniophotomoeter in place using the screws at the top. Once the position and level of the Goniophotomoeter main unit have been properly adjusted, do not move it again, as this will cause deviations in the test results.

5.2 광도계 프로브의 교정

5.2.1 0도 위치에서의 광도 측정 프로브 교정

번호 주의 사항
0 입력
회전 방향을 클릭하면 감마 축이 0도 위치로 되돌아갑니다.

High Precision Rotation Luminaire Goniophotometer-Figure34

그림 39

Verify that the single-line laser beam is passing through the exact horizontal center of the last partition at this point.

High Precision Rotation Luminaire Goniophotometer-Figure35

그림 40

그렇지 않다면 아래 단계를 따르십시오.

번호 주의 사항
실제 상황에 따라 1, 0.1, 0.05 등과 같은 작은 각도 값을 입력해 주세요.
Select a rotation direction and rotate the Gamma axis until the single-line laser passes through the exact center of the last partition horizontally.
Click “Set to 0 Degrees” to set the current position of the Gamma axis on the near-field luminance probe to 0 degrees. The password is LISUN그룹.

High Precision Rotation Luminaire Goniophotometer-Figure36

그림 40.1

참고 1: 근접장 광도계의 감마 축을 0도로 설정한 후에는 임의로 다시 0도로 설정하지 마십시오. 그렇게 하면 테스트 결과에 오차가 발생할 수 있습니다.

Note 2: Definition of the Gamma angle for a goniophotomoeter. The direction directly facing the photometer probe is defined as 0 degrees/360 degrees. When rotating clockwise, the angle value increases from 0 degrees to 360 degrees; when rotating counterclockwise, the angle value decreases from 360 degrees to 0 degrees. Since angle notation may vary, -90 degrees and 270 degrees correspond to the same position.

5.2.2 광도 측정 프로브 장착

Rotate the light sensor’s lens cover counterclockwise and remove it.

High Precision Rotation Luminaire Goniophotometer-Figure37

그림 41

The standard lamp packaging box contains a small mirror.

High Precision Rotation Luminaire Goniophotometer-Figure38

그림 42

High Precision Rotation Luminaire Goniophotometer-Figure39

그림 43

Remove the small reflector and attach it to the probe.

High Precision Rotation Luminaire Goniophotometer-Figure40

그림 44

At this point, the Gamma axis of the near-field photometer is still at the 0-degree position. Slightly move the probe base and adjust the screws at points ① and ② to adjust the probe’s height and orientation. Position the probe so that the straight laser beam coming from the main axis hits the exact center of the mirror, and ensure that the reflected beam travels back along the same path.

High Precision Rotation Luminaire Goniophotometer-Figure41

그림 45

그런 다음 프로브 브래킷 받침대를 나사를 사용하여 바닥에 고정합니다.

High Precision Rotation Luminaire Goniophotometer-Figure42

그림 46

Once secured, you may need to make slight adjustments to the probe’s height and orientation so that the straight laser beam continues to hit the exact center of the reflector and the reflected beam returns along the same path. After calibration is complete, ensure that all screws are tightened. Then remove the reflector and reinstall the light probe’s light shield.

Note: Once the probe’s position and orientation have been set, do not adjust them, as doing so may cause the test data to be inaccurate.

You can secure the photometer probe’s power adapter to the probe mount as shown in the figure below to protect the probe’s power jack.

High Precision Rotation Luminaire Goniophotometer-Figure43

그림 47

5.3 크로스라인 레이저 설치

마지막으로, 십자선 레이저는 암실 측벽에 설치해야 합니다. 각 암실의 구조에 따라 십자선 레이저는 좌측 또는 우측 벽에 설치할 수 있습니다. 아래 그림에 나타난 암실 구조를 예로 들어 설명하겠습니다.

High Precision Rotation Luminaire Goniophotometer-Figure44

그림 52

각도 조절 화면으로 이동하세요.

번호 주의 사항
90을 입력하십시오 (암실 설계도에 표시된 레이저 조준선의 설치 위치가 다를 경우 -90을 입력하십시오).
Click to rotate the gamma axis to a 90-degree position, so that the main-axis laser is directly aligned with the position where the crosshair laser is installed.

High Precision Rotation Luminaire Goniophotometer-Figure45

그림 53

Secure the crosshair laser to the triangular mount we provide.

High Precision Rotation Luminaire Goniophotometer-Figure46

그림 54

Turn on the crosshair laser and use the direction of the single-line laser to roughly determine the position of the crosshair laser.

High Precision Rotation Luminaire Goniophotometer-Figure47

그림 55

Fine-Tune the Position of the Crosshair Laser: The centerline of the straight laser should be perfectly aligned with the centerline of the crosshair laser. You can verify and adjust the position using the following method. Place a sheet of white paper at the exit of the cross-shaped laser and at the exit of the single-line laser. If the centerline of the single-line laser aligns with the centerline of the cross-shaped laser at both locations, you can confirm that the position and orientation of the cross-shaped laser are correct.

High Precision Rotation Luminaire Goniophotometer-Figure48

그림 56

High Precision Rotation Luminaire Goniophotometer-Figure49

그림 57

Next, secure the cross-laser tripod to the wall using screws. Once secured, you can still loosen the screws if necessary to make slight adjustments to the cross-laser in both the horizontal and vertical directions, ensuring it remains properly aligned. After calibration is complete, all screws should be tightened securely.

High Precision Rotation Luminaire Goniophotometer-Figure50

그림 58

Then rotate the Gamma axis back to the 0-degree position. The cross-shaped laser beam is projected onto the opposite wall. You can use this to create a permanent mark. That is, when the cross-shaped laser beam is projected in this direction, it indicates the horizontal and vertical center positions of the Goniophotomoeter. This mark is needed for both calibration and testing.

High Precision Rotation Luminaire Goniophotometer-Figure51

그림 59

Installation and commissioning are complete. Turn off the line laser on the main shaft of the Goniophotomoeter.

5.4 청소

You may shut down all equipment. When doing so, first turn off the power switch on each individual device, and then turn off the main power switch for the cabinet. Next, thoroughly clean the darkroom and control room. To prevent dust from entering the equipment, cover the devices with dust-proof cloths before cleaning. After cleaning, turn all equipment back on and continue with the calibration and testing procedures.

참고: 암실을 사용하는 동안 먼지가 최대한 없도록 유지하십시오.

6. 보정

6.1 표준 램프의 설치 및 시운전

6.1.1 스탠드 조명 설치

First, turn on the crosshair laser on the wall and rotate it to ensure that it is aimed at the mark on the opposite wall.

스탠드 램프 상자를 여세요.

High Precision Rotation Luminaire Goniophotometer-Figure52

그림 60

다음으로 아래 그림과 같이 표준 램프 브래킷을 조립하십시오.

High Precision Rotation Luminaire Goniophotometer-Figure53

그림 61

The main shaft of the Goniophotomoeter should still be aligned with the photometric probe, i.e., the probe’s 0-degree position. Locate the calibration rod and install it at the center of the Goniophotomoeter. Then remove the top section.

High Precision Rotation Luminaire Goniophotometer-Figure54

그림 62

아래 지침에 따라 전선을 연결해 주십시오.

번호 주의 사항
The main shaft of the goniophotomoeter has four terminals for connecting the power cables and sampling cables of the standard lamp and sample luminaire. Use short wires to bridge the red and yellow terminals together, and the black and blue terminals together.
Connect the power cord of the standard lamp/sample fixture to the red and blue terminals on the main shaft.

High Precision Rotation Luminaire Goniophotometer-Figure55

그림 63

6.1.2 표준 광 방향 교정

번호 주의 사항
스탠드 조명의 높이를 조절하려면 여기 있는 나사를 풀어주세요.
여기에는 스탠드 조명의 각도를 조절할 수 있는 나사가 세 개 있습니다.

High Precision Rotation Luminaire Goniophotometer-Figure56

그림 64

The standard lamp should be adjusted so that the cross-shaped laser beam is directed at the center of the small mirror, and the horizontal and vertical laser beams reflected from the mirror coincide exactly with the original cross-shaped laser beam.

High Precision Rotation Luminaire Goniophotometer-Figure57

그림 65

Based on our experience, follow these steps for calibration: First, adjust the height of the standard lamp to the appropriate position. The vertical reflected laser can be calibrated by rotating the calibration target, while the horizontal reflected laser can be calibrated by adjusting the pitch of the standard lamp. After calibration, the center height of the standard lamp may change, requiring you to readjust its height. Continue the reflection adjustment until the cross-shaped laser beam is projected onto the center of the small mirror, and the reflected horizontal and vertical laser beams perfectly align with the original cross-shaped laser beam.

Once the standard lamp is properly aligned, carefully remove the small reflector, taking care not to alter the alignment of the standard lamp during removal. Then turn off the cross-line laser.

6.2 광도계 프로브의 교정

All standard lamps are constant-current DC light sources. Set the cabinet’s AC/DC switch to DC.

High Precision Rotation Luminaire Goniophotometer-Figure58

그림 66

소프트웨어 각도 제어기를 엽니다.

번호 주의 사항
Enter -90
감마 축을 -90도 위치로 회전시키려면 클릭하십시오(암실 도면에서 십자선 레이저가 다른 위치에 설치된 경우 90도 위치로 회전시켜야 합니다).

High Precision Rotation Luminaire Goniophotometer-Figure59

그림 66.1

표준 램프 마운트의 둥근 구멍이 광도 측정 프로브를 향하도록 정렬하십시오.

High Precision Rotation Luminaire Goniophotometer-Figure60

그림 67

교정 램프를 켜고 교정 인터페이스로 들어갑니다.

번호 주의 사항
Select a DC Power Supply
표준 램프는 정전류 광원입니다. CC(정전류 모드)를 선택하고 표준 램프 인증서에 명시된 기준 전압과 정격 전류를 입력하십시오. 일반적으로 이 값은 30V 및 6.1A입니다.
클릭하여 표준 램프를 켜고 암실의 다른 모든 조명을 끄십시오. 표준 램프가 안정적인 상태에 도달하는 데 15분이 소요됩니다.
표준 램프가 안정화되면 클릭하여 광도계 교정 인터페이스로 들어갑니다.

High Precision Rotation Luminaire Goniophotometer-Figure61

그림 68

암실 문을 닫고 스탠드 조명을 제외한 암실 내 모든 조명이 꺼져 있는지 확인하십시오.

번호 주의 사항
근접장 광도 측정 프로브 선택하기
Luminous flux calibration—our software includes this option, but current standard practices require the use of luminous intensity calibration, and the standard lamps included in our standard configuration are also luminous intensity standard lamps. Therefore, there is no need to use luminous flux calibration at this time.
Light Intensity Calibration Method
Refer to the standard lamp calibration certificate and enter its calibrated luminous intensity value.
시작하려면 클릭하세요
소프트웨어가 테스트 거리를 자동으로 계산합니다.
결과를 확인하려면 "확인"을 클릭하십시오. 그러면 소프트웨어에 현재 측정된 광량 값이 표시됩니다. 측정된 광량 값은 보정된 광량 값과 일치해야 합니다.
저장하고 종료하려면 "확인"을 클릭하세요.

High Precision Rotation Luminaire Goniophotometer-Figure62

그림 69

참고: 측정 거리는 고니오포토미터의 중심에서 광도계 프로브 표면까지의 거리입니다. 소프트웨어에서 계산된 거리는 실제 거리와 거의 일치해야 하며, 1% 이하의 오차는 정상 범위로 간주됩니다. 오차가 너무 큰 경우, 2.3.1절을 참조하여 암실이 정상적으로 작동하는지 확인하고, 5장을 참조하여 교정 단계를 검토하십시오.

교정이 완료되면 소프트웨어를 사용하여 표준 램프를 끌 수 있습니다. 표준 램프는 사용 중에 상당히 뜨거워지므로, 안전한 보관을 위해 포장재에 다시 넣기 전에 완전히 식을 때까지 기다리십시오. 교정봉을 분리하여 안전하게 보관하십시오.

6.3 교정 및 유지 관리

암실에는 아무런 변화가 없습니다. 컴퓨터에 변경 사항이 없다면 재보정은 필요하지 않습니다.
No changes were made to the computer; simply unplug and replug the communication cable or reinstall the Goniophotomoeter software. No recalibration is necessary.
If you switch to a new computer or reinstall the operating system, there is no need to recalibrate; simply enter the previous test distance and save it.
Changes in the Darkroom 암실의 구조는 변하지 않지만, 장기간 사용 후에는 내부 페인트가 바래거나 벗겨질 수 있습니다. 암실 관리 상태와 테스트 결과의 정확도에 따라 주기적으로 암실을 다시 칠해야 할 수도 있습니다. 재도색 후에는 암실을 재보정해야 합니다.
암실 구성이 변경되어 재보정이 필요합니다.

7. Preparations Before Testing

7.1 Lamp Selection and Aging

Select representative luminaires for testing that comply with national or corporate standards; new luminaires must undergo aging treatment (for LED luminaires, a 24-hour aging period is recommended);

조명기구에 눈에 띄는 손상이 없는지, 내부 배선이 제대로 연결되어 있는지 확인하고, 조명기구 표면의 얼룩을 깨끗이 닦아내십시오.

7.2 테스트 모드 소개

C-Gamma Mode: Suitable for most lighting fixtures, such as streetlights, downlights, panel lights, and plant grow lights.

High Precision Rotation Luminaire Goniophotometer-Figure63

그림 74

베타 모드: 스포트라이트, 투광등 및 유사 조명기구에 적합하며, 보조 암 설치가 필요합니다.

High Precision Rotation Luminaire Goniophotometer-Figure64

그림 75

알파 모드: 자동차 조명 및 방향지시등에 ​​적합합니다.

High Precision Rotation Luminaire Goniophotometer-Figure65

그림 76

참고 : LSG-1890BCCD can perform C-Gamma and B-Beta tests. For A-Alpha testing, a Type A goniophotomoeter is required; please contact us if needed.

8. C-감마 테스트

8.1 조명 설치 및 켜기

8.1.1 Light Fixture Installation

저희는 두 가지 C-감마 테스트 픽스처를 보유하고 있으며, 이 절차에서는 십자형 픽스처를 예시로 사용합니다.

번호 주의 사항
크기가 큰 십자형 클램프는 다양한 종류의 조명 기구에 사용됩니다.
E27 소켓을 이용한 구형 전구 테스트용

High Precision Rotation Luminaire Goniophotometer-Figure66

그림 77

Attach the fixture to the Goniophotomoeter. Turn on the wall-mounted cross-line laser and verify that the cross-line laser is projected accurately onto the markings made on the opposite wall.

High Precision Rotation Luminaire Goniophotometer-Figure67

그림 78

On the angle control interface, select the near-field photometric probe. Rotate the Gamma axis 90 degrees so that the crosshair laser points directly at the center of the crosshair fixture. Mount the lamp under test onto the crosshair fixture. Adjust the lamp’s position so that the crosshair laser is centered on the light-emitting surface of the lamp under test, and the light-emitting surface is parallel to the crosshair fixture.

High Precision Rotation Luminaire Goniophotometer-Figure68

그림 79

그런 다음 감마 축을 0도로 되돌립니다. 여기에서 다이얼을 돌리세요.

High Precision Rotation Luminaire Goniophotometer-Figure69

그림 80

Align the light-emitting surface of the fixture with the vertical laser beam of the crosshair.

High Precision Rotation Luminaire Goniophotometer-Figure70

그림 81

Caution is required when installing directional luminaires. When the angle of the C-plane on the Goniophotomoeter is 0 degrees, according to streetlight installation standards, the light pole should point directly upward, as shown in the figure below.

High Precision Rotation Luminaire Goniophotometer-Figure71

그림 81.1

고니오포토미터의 C면 각도가 0도일 때, 스트립 조명은 아래 그림과 같이 업계 표준에 따라 수직으로 설치해야 합니다.

High Precision Rotation Luminaire Goniophotometer-Figure72

그림 81.2

8.1.1 조명 켜기

테스트 대상 조명기구에 전원 코드를 연결하십시오. 테스트 대상 조명기구의 연결 방법은 표준 조명기구의 연결 방법과 동일합니다. 6.1.1절을 참조하십시오.

테스트 중인 조명기구의 매개변수에 따라 캐비닛 스위치를 조정하십시오.

번호 주의 사항
If the sample luminaire is AC-powered, switch to the AC setting; if it is DC-powered, switch to the DC setting.
전기 파라미터용 내부/외부 샘플링 스위치입니다. 외부 샘플링은 4선식 구성을 사용하여 보다 정확한 전압 측정값을 제공합니다. 별도의 설정이 없는 한, 이 스위치를 "외부 샘플링"으로 설정하십시오. 추가 조정은 필요하지 않습니다.

High Precision Rotation Luminaire Goniophotometer-Figure73

그림 82

샘플 조명기구가 교류 전원을 사용하는 경우.

번호 주의 사항
Select AC Power
샘플 매개변수를 기준으로 올바른 전압과 주파수를 입력하십시오.

High Precision Rotation Luminaire Goniophotometer-Figure74

그림 83

샘플 조명기구가 직류 전원인 경우.

번호 주의 사항
Select a DC Power Supply
샘플 매개변수를 기준으로 올바른 전압과 전류를 입력하십시오.
Select the output mode: CC stands for constant current, and CV stands for constant voltage. If you select constant current, you need to enter the sample’s rated current value; for the voltage, simply enter the maximum output voltage of this DC power supply. If you select constant voltage, enter the sample’s rated voltage and the DC power supply’s maximum output current.
참고: DC 전원 공급 장치 모델이 다음과 같은 경우 DC3010, its maximum output voltage is 30V and its maximum output current is 10A.

High Precision Rotation Luminaire Goniophotometer-Figure75

그림 84

The software offers two methods for turning the sample lamp on and off, and users can choose between them.

번호 주의 사항
If you check the “Auto Turn On” and “Auto Turn Off” options, the software will automatically power on the sample according to the selected power source and the entered electrical parameters when the test begins, and will automatically cut off the sample’s power when the test ends.
If the “Auto On” and “Auto Off” options are not checked, you can still click this button; the software will illuminate the sample based on the selected power source and the entered electrical parameters. Clicking this button again will cut off the power to the sample.

High Precision Rotation Luminaire Goniophotometer-Figure76

그림 85

8.2 예열 매개변수 설정

번호 주의 사항
"예열" 확인란을 선택하면 소프트웨어는 테스트 시작 시 지정된 예열 매개변수에 따라 조명 기구를 예열하고, 예열이 완료되면 테스트가 자동으로 시작됩니다.
If you do not check the “Preheat” box, the software will start the test immediately.
예열 매개변수 설정

High Precision Rotation Luminaire Goniophotometer-Figure77

그림 86

Click “Preheat Parameter Settings.”

번호 주의 사항
예열 시간 설정: 예열의 목적은 조명 기구의 밝기와 색상 매개변수가 안정화될 때까지 일정 시간 동안 조명을 켜둔 상태를 유지하는 것이며, 이후 테스트가 자동으로 진행됩니다.
Sampling interval: The software automatically samples data at the set interval throughout the entire warm-up period to obtain real-time data. We recommend setting this to 5 seconds or longer.
If checked, the test will automatically skip the remaining warm-up time and proceed once the light fixture has reached a stable state; if unchecked, the test will begin only after the warm-up period has been completed according to the set duration.
Definition of a luminaire’s steady-state condition. If the values are set to 20 and 0.5%, the luminaire is considered to be in a steady state if the parameter has not varied by more than 0.5% over the past 20 minutes.
조명기구의 안정성을 판단하는 데 소프트웨어가 사용할 매개변수를 선택할 수 있습니다. 광학적 매개변수, 전기적 매개변수 등이 있습니다.
The software will save the change curves for the selected parameters during the warm-up period in the test report.
설정을 완료했으면 확인을 클릭하여 저장하고 종료하십시오.

High Precision Rotation Luminaire Goniophotometer-Figure78

그림 87

8.3 테스트 매개변수 구성 및 테스트 시작

Click “Start Test.”

High Precision Rotation Luminaire Goniophotometer-Figure79

그림 88

Go to the test parameter settings screen.

번호 주의 사항
Lighting fixture types: indoor lights, outdoor lights, or plant grow lights—these three types of lighting fixtures are subject to C-Gamma testing.
Select whether the luminaire is symmetrical or asymmetrical. Unless there are specific requirements, select the first option (asymmetrical) regardless of whether the luminaire being tested is symmetrical or not (to obtain the luminaire’s actual data).
C-plane angle range: Select 0 to 180 degrees; no modification is needed. For data ranging from 180 to 360 degrees, the software will automatically convert the
C-plane angle intervals. Symmetrical luminaires, such as indoor lights, can use a larger interval (e.g., 30 degrees), while asymmetrical luminaires, such as streetlights, can use a smaller interval (e.g., 10 degrees). The smaller the angle interval, the higher the test accuracy, but the longer the test will take. If time permits, we recommend selecting a smaller angle interval.
Gamma plane angle range: If the sample’s light irradiation angle exceeds 180 degrees (e.g., incandescent bulbs), select -180 degrees to 180 degrees; if the sample’s light irradiation angle is less than 180 degrees (e.g., panel lights), you may select a Gamma
plane angle interval ranging from -90 degrees to 90 degrees. Unless there are special requirements, a 1-degree interval is sufficient. The specific angle interval selected here has no effect on test speed.
If the system has multiple photometric probes, please select the one you want to use.
For standard luminaires undergoing standard testing, select “Normal Speed”; if the luminaire being tested has an extremely narrow beam angle, select “Slow Speed” to obtain more accurate test results; for expedited testing, select “Fast Speed.”
Fixed Range: If you are consistently measuring the same type of light fixture and the luminous intensity data does not vary significantly, you can check this box to save time. If you have ample time, or if you frequently test lamps with different luminous intensities, we recommend leaving this box unchecked; in this case, the photometric probe will automatically select the appropriate range for each lamp and enter
continuous scanning mode. This is suitable when the angular interval of the Gamma plane is between -180 degrees and 180 degrees. During the test, when the Gamma angle approaches ±180 degrees, simultaneously rotating the C-axis to the next scanning position—that is, the Gamma axis rotates continuously without stopping to save time. If you have ample time, we recommend leaving this option unchecked.
To set the monitoring angle during lamp warm-up, simply set it to (0, 0).
Normalization refers to the decision of whether to apply a uniform value to the first row of data in the luminous intensity table of a test report; certain standards, such as the U. S. streetlight standard, may require you to perform a similar adjustment. There are three options: no normalization, normalization to the maximum value (which will slightly increase the luminaire’s total luminous flux), and normalization to the average value (which leaves the luminous flux essentially unchanged).
If you need to remove all stray light outside a specific angle, check this box and enter the desired angle
. For example, if you check this option and enter 60, the data outside ±60 degrees of the luminaire’s gamma plane will automatically be set to 0.
Enter the number of light sources in the luminaire and the rated luminous flux of each light source. If you do not know these values, you may leave them blank; this will not affect the test results.
샘플의 발광 표면 치수를 입력하십시오. 실내 조명 기구의 경우, 소프트웨어는 휘도 한계 곡선, UGR 및 기타 매개변수를 계산하기 위해 이 데이터가 필요합니다.
Enter information about the sample light fixture
시험 환경(예: 암실)의 온도와 습도를 입력하십시오.
Click OK to go to the warm-up screen, or go to the test screen.

High Precision Rotation Luminaire Goniophotometer-Figure80

그림 89

Then wait for the software testing to finish.

9. 베타 테스트

9.1 Install and Turn On the Lights

9.1.1 Lighting Fixture Installation

십자선 레이저를 켜고 반대쪽 벽에 표시해 둔 부분을 레이저가 계속 비추고 있는지 확인하세요.

정사각형 B-Beta 테스트 픽스처를 찾으십시오.

High Precision Rotation Luminaire Goniophotometer-Figure81

그림 92

B-Beta 테스트를 수행할 때는 고니오포토미터의 보조 암을 다시 부착하십시오. 그런 다음 그림과 같이 B-Beta 고정 장치를 수직으로 설치하십시오.

High Precision Rotation Luminaire Goniophotometer-Figure82

그림 93

번호 주의 사항
Click to switch to the angle display mode
Click ① to display the B-Beta angle here instead of the C-Gamma angle.
클릭하여 각도 조절 인터페이스를 엽니다.

High Precision Rotation Luminaire Goniophotometer-Figure83

그림 94

번호 주의 사항
각도 B가 0도일 때 조명기구는 수직 위치에 있어야 합니다. 이는 수평계를 사용하여 확인할 수 있습니다.
0 입력
Click to set the Beta angle to 0 degrees. Note: The 0-degree position for the Gamma angle is not the same as the 0-degree position for the Beta angle.

High Precision Rotation Luminaire Goniophotometer-Figure84

그림 95

After rotating Beta to the 0-degree position, mount the floodlight onto the fixture, ensuring that the light-emitting surface of the floodlight faces the photometric probe. Adjust the mounting position of the fixture so that the cross-shaped vertical laser beam aligns with the light-emitting surface of the floodlight.

High Precision Rotation Luminaire Goniophotometer-Figure85

그림 96

To ensure the sample is centered in the Goniophotomoeter, rotate the Beta fixture to the position shown in the figure below so that the sample’s light-emitting surface faces the crosshairs. Use the crosshairs to fine-tune the position of the sample holder.

High Precision Rotation Luminaire Goniophotometer-Figure86

그림 97

Once the light fixture is installed, you can turn off the crosshair laser.

9.1.2 조명 켜기

8.1.2항을 참조하십시오.

9.2 예열 매개변수 설정

8.2절을 참조하십시오.

9.3 테스트 매개변수 구성 및 테스트 시작

Click “Start Test” to go to the test parameter settings screen.

번호 주의 사항
"투광 조명"을 선택하시면 테스트 모드가 자동으로 B-베타 모드로 전환됩니다.
B면의 각도 범위는 -90도에서 90도까지입니다. 조정할 필요가 없습니다.
평면 B의 각도 간격을 설정할 수 있습니다. 5도 또는 10도와 같이 더 작은 각도 간격을 선택할 수 있습니다. 간격이 작을수록 테스트 정확도는 높아지지만 시간이 더 오래 걸립니다.
The Beta plane angle range is -90 degrees to 90 degrees. There is no need to adjust the
Beta plane interval angle; unless there are special requirements, select 1 degree. Different interval angles here have no effect on test speed.

High Precision Rotation Luminaire Goniophotometer-Figure87

그림 98

For other parameter settings, refer to Section 8.1.3. Once you have configured the settings, click “Start” and wait for the software to complete the test.

10. 시험 보고서 처리

10.1 데이터베이스 파일 구성

번호 주의 사항
기본적으로 모든 테스트 보고서는 테스트 실행 순서대로 현재 열려 있는 데이터베이스 파일에 자동으로 저장됩니다.
보고서를 삭제하려면 해당 보고서를 선택한 다음 "삭제"를 클릭하십시오.
To make it easier to find test reports, click the blank button shown in the image below to create a new database file, and store reports for different lighting fixtures in separate database files. After creating a new database file, you can proceed directly with testing without having to recalibrate or reconfigure the system.
다른 데이터베이스 파일의 테스트 보고서를 보려면 "열기"를 클릭하십시오.

High Precision Rotation Luminaire Goniophotometer-Figure88

그림 100

Note 1: The next time the software is launched, it will automatically load the database file used during the previous session. If the path to that database file has changed or its name has been modified, the software will prompt you to create a new database file.

참고 2: 가능하면 데이터베이스 파일을 컴퓨터 바탕화면에 두지 마십시오.

10.2 테스트 보고서 표시 설정

번호 주의 사항
You can double-click to upload a picture of the light fixture.
이 영역에 표시된 내용을 두 번 클릭하면 편집할 수 있습니다.

High Precision Rotation Luminaire Goniophotometer-Figure89

그림 101

번호 주의 사항
"설정"을 클릭하십시오
단위 선택 시 일반적으로 "cd"가 사용되지만, 에너지 절약형 전구와 관련된 일부 구형 표준에서는 "cd/klm" 사용을 요구할 수 있습니다.
가로등 C0 각도 설정. CIE/GB 및 IESNA 표준은 요구 사항이 일관되지 않으므로, 사용하시는 표준을 선택하십시오.
Customize the divergence angle. The software automatically displays the half-peak angle (commonly referred to as the beam angle, calculated as 50% of the maximum luminous intensity) and the effective beam angle (calculated as 10% of the maximum luminous intensity). If you wish to customize the divergence angle, enter the percentage of the maximum light intensity and check the box (e.g., 0.25 for 25%). In this case, the test report will display not only the half-peak angle and effective beam angle but also your custom divergence angle.
보고서에 최대 광도 원뿔 각도를 표시하려면 이 확인란을 선택하십시오.

High Precision Rotation Luminaire Goniophotometer-Figure90

그림 102

번호 주의 사항
Switch to the “Energy Efficiency Rating” settings screen, then select the applicable standard and luminaire type.
현재 테스트 보고서에 적용하려면 클릭하세요.
Click OK to save and exit.

High Precision Rotation Luminaire Goniophotometer-Figure91

그림 103

Each display window has several settings that allow you to customize the display. Users can experiment with adjusting these settings to achieve their desired display effects.

High Precision Rotation Luminaire Goniophotometer-Figure92

그림 103.1

10.3 테스트 데이터 처리

10.3.1 Double-sided luminaires

For lamps that emit light from both sides, use the indoor lamp test mode to test each side individually—that is, set the gamma plane angle range to -90 degrees to 90 degrees during testing—and then generate two test reports.

번호 주의 사항
Hold down the “Ctrl” key on your computer keyboard, then click to select the two test reports you just received.
소프트웨어 탐색 모음에서 "데이터 합성"을 클릭하십시오.
In the pop-up window, you can click to toggle between the upward and downward light beams.
"복합"을 선택하세요.
저장하고 종료하려면 확인을 클릭하세요.

High Precision Rotation Luminaire Goniophotometer-Figure93

그림 104

이렇게 하면 아래 그림과 같이 새로운 합성 테스트 보고서가 생성됩니다.

High Precision Rotation Luminaire Goniophotometer-Figure94

그림 105

10.3.2 테스트 보고서 유형 변환

번호 주의 사항
“유형 변환”을 클릭하세요.
Test reports can be converted among four types
When converting to a C-Gamma test report, you can use the “termination angle” of the C-plane to generate a symmetric test report. If the termination angle is set to 360 degrees, there is no symmetry; if the termination angle is set to 180 degrees, the report can be converted to an axially symmetric test report; if the termination angle is set to 90 degrees, the report can be converted to a quadrant-symmetric test report; if the termination angle is set to 0 degrees, the report can be converted to a fully symmetric test report. This feature is primarily intended for symmetric indoor lighting fixtures; certain lighting design software programs, such as DIALUX, may require you to use this setting when importing fully symmetric IES files.

High Precision Rotation Luminaire Goniophotometer-Figure95

그림 106

10.4 Exporting Test Data and Printing Test Reports

10.4.1 테스트 데이터 내보내기

번호 주의 사항
Click the red logo in the upper-left corner of the software
드롭다운 메뉴에서 "데이터 내보내기"를 클릭하세요.

High Precision Rotation Luminaire Goniophotometer-Figure96

그림 107

Files can be exported in a variety of standard formats.

High Precision Rotation Luminaire Goniophotometer-Figure97

그림 108

10.4.2 인쇄 테스트 보고서

인쇄 설정.

번호 주의 사항
Click to open the “Print Settings” screen
The report displays language options; you can choose between Chinese and English.
페이지 번호 형식
Report Header Information
보고서 제목 (중국어 및 영어)
General Report Content—Applicable to Reports on All Types of Lighting Fixtures
Since indoor lights, streetlights, floodlights, and plant grow lights have different key parameters, the report content will vary accordingly. You can select the specific items you want to include in the report.
TM-30 옵션
CIE color space diagram or SDCM color tolerance diagram—you can choose only one of them.

High Precision Rotation Luminaire Goniophotometer-Figure98

그림 109

번호 주의 사항
현재 선택된 테스트 보고서를 인쇄하려면 클릭하십시오.
Click to print all test reports in the current database file.

High Precision Rotation Luminaire Goniophotometer-Figure99

그림 110

11. 장비의 일상 유지보수 및 관리

11.1 일상적인 유지보수

시작 절차: 먼저 캐비닛의 주 전원 스위치를 켜십시오. 그런 다음 캐비닛 내부 및 암실에 있는 장비의 전원 스위치를 하나씩 켜십시오. 마지막으로 소프트웨어를 실행하십시오. 종료 절차: 먼저 소프트웨어를 종료하십시오. 그런 다음 캐비닛 내부 및 암실에 있는 장비의 전원 스위치를 하나씩 끄십시오. 마지막으로 캐비닛의 주 전원 스위치를 끄십시오.

Keep the darkroom control room clean and dust-free to prevent dust from entering the equipment. If there is surface dust on the equipment, wipe it off with a dry, soft cloth.

Avoid exposing the darkroom to high humidity for extended periods; keep a dehumidifier on hand if necessary.

11.2 일상적인 유지보수

11.2.1 광도 측정 프로브 청소

If the darkroom cannot be kept dust-free, clean the light-sensitive surface of the photometer probe regularly (for example, once a month) based on actual conditions to prevent dust from accumulating and affecting test accuracy.

When cleaning, please use a professional lint-free cloth to remove dust without scratching the device.

11.2.2 Standard Lamp

표준 램프의 유효 기간은 1년입니다. 1년 후에도 암실 교정이 필요한 경우 새 표준 램프를 구매하시는 것을 권장합니다.

11.2.3 소프트웨어

저희는 기능 추가 및 버그 수정을 위해 소프트웨어를 정기적으로 업데이트할 예정입니다. 소프트웨어 사용 중 문제가 발생하면 언제든지 저희에게 연락하여 최신 버전을 받으시기 바랍니다.

11.3 장기 해체 및 유지보수

11.3.1 장비 청소 및 보호

모든 장비 표면을 철저히 청소하고 본체, 프로브 및 케이스를 방진 천으로 덮은 후 모든 전원을 차단하십시오.

11.3.2 환경 보호

Ensure that the darkroom is dry and well-ventilated to prevent equipment from getting damp. If the ambient humidity is high, keep a dehumidifier on hand.

11.3.3 Monthly Power-On

매달 전원 켜짐 테스트를 실시하여 모든 시스템 하드웨어가 제대로 작동하는지 확인하십시오.

12. 문제 해결 및 일반적인 문제 해결 방법

12.1 통신 오류

Cause of the problem: Check whether the communication cable is securely connected and whether it is plugged into the wrong port; verify that the driver is installed correctly; try a different USB port or communication cable to rule out hardware issues.

Solution: Reinstall the driver, connect the communication cable according to the device identifier, restart the software, and try again.

12.2 테스트 오류

12.2.1 Luminous Flux Anomalies

Cause of the malfunction: Check whether the light source is stable and whether the warm-up time is sufficient; verify that the equipment calibration is valid, that the test distance is correct, and that the standard light is functioning properly; check whether there is any external light interference in the darkroom and whether the light path is obstructed.

해결 방법: 예열 시간을 늘리고, 기기를 재보정하고, 암실의 다른 광원을 끄고, 광 경로의 장애물을 제거하십시오.

12.2.2 UGR 이상 현상

오류 원인: 조명기구의 발광면 치수를 입력하지 않았습니다. 조명기구의 빔 각도가 45° 미만입니다. 계산 기준이 잘못 선택되었습니다.

해결 방법: 조명기구의 발광면 치수(길이/너비/높이)를 정확하게 입력하십시오. 조명기구가 실내용인지 확인하십시오(빔 각도가 충분해야 함). CIE 190:2010 표준을 선택하고 다시 계산하십시오.

12.3 고니오포토미터는 회전하지 않습니다

Cause of the malfunction: Check whether the host power is on and whether the wiring is correct; check whether the angle controller inside the cabinet is operating normally; confirm whether any foreign objects are jammed in the rotating shaft.

Solution: Check the power supply and wiring, and remove any foreign objects; restart the system and software, then try again.

12.4 Software Malfunctions

12.4.1 소프트웨어 충돌

문제 원인: 관리자 계정으로 컴퓨터 운영 체제에 로그인했는지 확인하십시오. 소프트웨어가 완전히 설치되었는지 확인하십시오. 데이터베이스 파일이 손상되었는지 확인하십시오.

Solution: Log in to the computer using an administrator account; reinstall the software and drivers; create a new database file.

12.4.2 소프트웨어에 "파일이 데이터베이스가 아닙니다"라는 메시지가 표시됩니다.

문제 발생 원인: 데이터베이스 파일이 삭제되었거나 이동되었습니다. 또는 LGF 파일을 여는 데 사용되는 기본 프로그램이 변경되었습니다.

Solution: Create a new database file and save it; restore the default association for LGF files.

12.5 정전

12.5.1 직류 전원 출력 이상 현상

Causes of Failure: The cabinet’s AC/DC switch was not set to DC; the sample voltage or current exceeded the DC power supply limits; a short circuit in the sample;

문제 해결: 샘플의 전압과 전류가 DC 전원 공급 장치의 출력 범위 내에 있는지 확인하십시오. 예를 들어, DC3005 해당 모델의 최대 출력 전압은 30V이고 최대 출력 전류는 5A입니다. 샘플에 단락이 있는지 확인하고, 단락이 있는 경우 재시험 전에 샘플을 수리하십시오.

12.5.2 AC 출력 없음

Cause of the fault: The cabinet’s AC/DC switch was not set to the AC position; the overload protection was triggered;

Troubleshooting Steps: Check the AC power output wiring and sample connections to ensure there are no loose connections or short circuits; if the overload protection is triggered, turn off the power, allow the unit to cool for 5 minutes, and then restart it. Ensure that the sample power remains within the AC power output range; for example, the LSP-500VARC has a maximum output power of 500W.