자체 안정 부하용 스위치 테스트
사용자 설명서
적용 할 수있는 LISUN 모델: SBLL-3P20A
Please open all packages and inspect the product according to the instructions in this chapter. If you notice any abnormalities, please contact us promptly.
This equipment is designed and manufactured in accordance with Clause 19.3 of Standard GB16915.1-2014 and Clause 19.3 of IEC 60669-1:2017.
Equipment front panel. = 1 * GB3① Power meter; 3 power meters for 3 workstations; = 2 * GB3② Touchscreen; = 3 * GB3③ Emergency stop; = 4 * GB3④ Alarm; = 5 * GB3⑤ Power switch;= 6 * GB3⑥: 12 terminals in total for 3 workstations; for single-pole switches, short N to L and connect L to the switch; for double-pole switches, connect both L and N to the switch;= 7 × GB3⑦ Signal terminals: Output signals when an alarm is triggered at a corresponding workstation. Pin 1 corresponds to Workstation 1, Pin 2 to Workstation 2, Pin 3 to Workstation 3, and Pin 4 is the common terminal (output signal is a digital signal). These are used to send signals to the switch life tester to control workstation operation;= 8 * GB3⑧ Measurement terminal blocks, used to connect an oscilloscope for equipment measurement.
Note: If = 7 * GB3⑦ and = 8 * GB3⑧ are not connected, it does not affect the operation of the load cabinet’s testing functions.

Accessories: Signal cable, power cable, ground cable, and workstation sample cable.

Touchscreen interface.
First row: Three-station display.
Second row: Test mode selection. Selected test modes are displayed in green; unselected test modes are displayed in red. To select a different mode for the same station, deselect the current mode and then select the desired mode.
Third row: Alarm delay time. When the device remains in the same state (closed or open) for the corresponding duration, an alarm will sound. When the device starts a test, this setting automatically changes from “Off” to “On” and begins timing; tap it manually to return it to the “Off” state.
Fourth row: Test pass/fail count.
Fifth row: When “Test” is displayed, the instrument is ready to perform normal testing. Click this button to switch the display to “Debug,” which enables metrological calibration of the device. Test start and end buttons.
Sixth row: Workstation status indicator.
Seventh row: “Test Voltage”—tap to enter the test voltage settings interface. “Alarm Set”—tap to enter the alarm delay time settings.

시험 절차
When connecting the device’s power cord, please verify that the input voltage is correct. When connecting the ground wire, please verify that the device is properly grounded.

For a three-station wiring setup, simply connect the wires to the number of stations required. Note: The figure below is for demonstration purposes only; in general, the wires should be connected to a switch life tester.

Turn on the device. Click “Alarm Set” and set the alarm delay time according to your needs.

Click “Test Voltage” to set the test voltage to the desired value (250V), then click “POWER ON.” The displayed voltage will gradually adjust automatically to match the set voltage.
Note: “Voltage compensation” generally does not need to be set. If the final displayed voltage does not match the set voltage, please contact us.

= 1 * GB3① Verify that the device is in “Test” mode, = 2 * GB3② Select the desired test setting, = 3 * GB3③ Click to start the test.

When the continuous “on” or “off” state reaches the set alarm delay time, the device will trigger an alarm, and testing for that setting will automatically stop. Click “On” in the image below to reset the device.

After the test is complete, click “Test Voltage,” then click “Power Off” to turn off the test voltage. Once the test voltage has automatically returned to 0, you may turn off the device’s power.
Debugging and Calibration Procedures
Note: Before adjusting any parameter, be sure to stop the test on the touchscreen first.
Connect the sample leads under normal test conditions.
Using a 1x probe, connect the oscilloscope as follows.

All settings are configured for normal test mode. Click “Test” to switch to “Debug” mode.

Using a single test station with a 10A test range as an example, click to start the test for the first test station.

Keep the sample in this station in the closed position at all times. The equipment will automatically apply and remove the surge current for calibration.
Oscilloscope settings: Trigger mode: Single; Probe multiplier: 1x (1X); Test current: Yes, 100 A/V. Capture the following waveforms and data: = 1 * GB3①Imax (standard value Ipeak is 108) and = 2 * GB3②I2t (standard value is 2.8), with a permissible error of ±5%.
Note: For data on other ranges, please also refer to Clause 19.3 of GB16915.1-2014 and Clause 19.3 of IEC60669-1:2017.

First, adjust the Ipeak surge current. As shown in the figure below, move the slider for the corresponding surge resistance setting to the left to decrease the surge resistance, which will increase the surge current. Conversely, move it to the right to increase the surge resistance, which will decrease the surge current.

After making the adjustments, use the oscilloscope to check the inrush current (Ipeak) again. If it is still incorrect, repeat the steps above.
After adjusting Ipeak, proceed to calibrate I2t. As shown in the figure below, you may need to increase or decrease the corresponding capacitance. The capacitance is pre-calibrated at the factory and generally does not require adjustment. If changes are needed, please contact us.

Calibrating the power resistor. = 1 * GB3① Set the device back to “Test” mode, = 2 * GB3② then click “Start Test,” = 3 * GB3③ Once the sample is closed, the corresponding power meter displays the power.
Locate the corresponding power resistor.
See the figure below. If the power is incorrect, loosen the clamp on the corresponding power resistor and slide it inward to reduce the resistance and increase the power; conversely, slide it outward to increase the resistance and decrease the power.



LED용 광도, 색도 및 안전 성능 테스트 솔루션
고니오포토미터 시스템: LSG-6000, LSG-1890B및 LSG-1800ACCD
분광복사계 및 적분구 테스트 시스템: LPCE-2 or LPCE-3
LED Life는 LM-84에 따라 테스트 시스템을 유지합니다. LEDLM-84PL
광생물학적 방사선 안전성 시험 시스템: EN 62471-C 또는 EN 62471-P
램프 시동, 예열 시간 및 깜빡임 테스트 시스템: LSRF-3 LSP-500VARC-Pst
LED 전원 드라이버 테스터: LS2090, M9822예산 및 LSP-500VARC
IK 레벨 테스터 | 스프링 해머 충격 시험기: IK07-10, IK01-06
전기역학적 진동 발생기 시스템: LVD-100KG-6D
전기 안전 검사원: LS9955, ZRS-3H, ZY-3, TTC-1
LED 등기구용 환경 챔버 테스트 솔루션
IPX5 및 IPX6 방수 테스트: JL-56
IP5X 및 IP6X 방진 테스트 장비: SC-015
High- and Low-Temperature and Humidity Chamber: GDJS-015B
염수 분무 시험기: YWX/Q-010
UV 램프 노화 시험 챔버: UV-263LS
크세논 램프 노화 테스트 챔버: XD-80LS
오존 시험 챔버: OTC-150A
이산화황 시험 챔버: SQ-010
CFL 및 LED 조명기구용 EMC 및 EMI 테스트 솔루션
EMI 테스트 시스템: EMI-9KB or EMI-9KA
정전기 방전 시뮬레이터: ESD61000-2
EFT 면역력 측정: EFT61000-4
서지 생성기: SG61000-5
전압 강하 및 정전 발생기: CSS61000-11
링 웨이브 발생기: RWG61000-12

WWW. Lisun그룹닷컴
특허 출원
모든 권리는 다음에 의해 보호됩니다. Lisun 그룹
사전 고지 없이 변경될 수 있습니다.

中文简体